Comparison of High-Resolution Fourier Transform Mass Spectrometry Platforms for Metabolite Annotation - C. elegans
Ontology highlight
ABSTRACT: Fourier transform ion cyclotron resonance (FT-ICR) and Orbitrap mass spectrometry (MS) are among the highest-performing analytical platforms in metabolomics. Their high mass measurement accuracy and mass resolving power enable detailed investigation of biological metabolomes. Non-targeted MS experiments, however, yield extremely complex datasets that make metabolite annotation very challenging, if not impossible. High-resolution accurate mass measurements greatly facilitate this process by reducing mass errors and spectral overlaps. When applied together with relative isotopic abundance (RIA) measurements, heuristic rules, and constraints during searches, the number of candidate elemental formula(s) can be significantly reduced. Here, we evaluate the performance of two leading analytical MS platforms, Orbitrap ID-X and 12T solariX FT-ICR mass spectrometers, in terms of mass accuracy and RIA measurements, and how these factors affect the assignment of the correct elemental formulae in metabolite annotation. Quality of the mass measurements was evaluated under various experimental conditions (resolution: 120 K, 240 K, 500 K; automatic gain control: 5e4, 1e5, 5e5) for the Orbitrap MS platform.