High-resolution imaging of the excised porcine heart at a whole-body 7 T MRI system using an 8Tx/16Rx pTx coil.


ABSTRACT:

Introduction

MRI of excised hearts at ultra-high field strengths ([Formula: see text]≥7 T) can provide high-resolution, high-fidelity ground truth data for biomedical studies, imaging science, and artificial intelligence. In this study, we demonstrate the capabilities of a custom-built, multiple-element transceiver array customized for high-resolution imaging of excised hearts.

Method

A dedicated 16-element transceiver loop array was implemented for operation in parallel transmit (pTx) mode (8Tx/16Rx) of a clinical whole-body 7 T MRI system. The initial adjustment of the array was performed using full-wave 3D-electromagnetic simulation with subsequent final fine-tuning on the bench.

Results

We report the results of testing the implemented array in tissue-mimicking liquid phantoms and excised porcine hearts. The array demonstrated high efficiency of parallel transmits characteristics enabling efficient pTX-based B1+-shimming.

Conclusion

The receive sensitivity and parallel imaging capability of the dedicated coil were superior to that of a commercial 1Tx/32Rx head coil in both SNR and T2*-mapping. The array was successfully tested to acquire ultra-high-resolution (0.1 × 0.1 × 0.8 mm voxel) images of post-infarction scar tissue. High-resolution (isotropic 1.6 mm3 voxel) diffusion tensor imaging-based tractography provided high-resolution information about normal myocardial fiber orientation.

SUBMITTER: Terekhov M 

PROVIDER: S-EPMC10140105 | biostudies-literature | 2023 Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

High-resolution imaging of the excised porcine heart at a whole-body 7 T MRI system using an 8Tx/16Rx pTx coil.

Terekhov Maxim M   Elabyad Ibrahim A IA   Lohr David D   Hofmann Ulrich U   Schreiber Laura M LM  

Magma (New York, N.Y.) 20230407 2


<h4>Introduction</h4>MRI of excised hearts at ultra-high field strengths ([Formula: see text]≥7 T) can provide high-resolution, high-fidelity ground truth data for biomedical studies, imaging science, and artificial intelligence. In this study, we demonstrate the capabilities of a custom-built, multiple-element transceiver array customized for high-resolution imaging of excised hearts.<h4>Method</h4>A dedicated 16-element transceiver loop array was implemented for operation in parallel transmit  ...[more]

Similar Datasets

| S-EPMC2861758 | biostudies-literature
| S-EPMC9622972 | biostudies-literature
| S-EPMC7116213 | biostudies-literature
| S-EPMC7536221 | biostudies-literature
| S-EPMC7924517 | biostudies-literature
| S-EPMC4732937 | biostudies-literature
| S-EPMC5681898 | biostudies-literature
| S-EPMC6617816 | biostudies-literature
| S-EPMC6466328 | biostudies-literature
| S-EPMC7496730 | biostudies-literature