Whole-body MRI using a sliding table and repositioning surface coil approach.


ABSTRACT:

Objective

To introduce and assess a new way of performing whole-body magnetic resonance imaging (MRI) using a non-integrated surface coil approach as available on most clinical MRI systems worldwide.

Methods

Ten consecutive asymptomatic subjects prospectively underwent whole-body MRI for health screening. Whole-body MRI included T1-, T2- and diffusion-weighted sequences, and was performed using a non-integrated surface coil to image four different stations without patient repositioning. The four separately acquired stations were merged, creating seamless coronal whole-body T1-, T2- and diffusion-weighted images. Anatomical alignment, image quality at the boundaries of adjacent stations, and overall image quality of all stations were qualitatively assessed.

Results

The average time (+/-SD) taken to change the surface coil from one station to the next station was 53.8 (+/-7.1) s. The average total extra examination time +/- SD was 2 min 41.4 s (+/-15.3 s). Anatomical alignment, image quality at the boundaries of adjacent stations, and overall image quality of all stations of T1-, T2- and diffusion-weighted whole-body MRI were overall graded as "good" to "excellent".

Conclusion

This study shows that a time-efficient and high-quality whole-body MRI examination can easily be performed by using a non-integrated sliding surface coil approach.

SUBMITTER: Takahara T 

PROVIDER: S-EPMC2861758 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC4732937 | biostudies-literature
| S-EPMC7716937 | biostudies-literature
| S-EPMC8663560 | biostudies-literature
| S-EPMC6996715 | biostudies-literature
| S-EPMC7771991 | biostudies-literature
| S-EPMC7555579 | biostudies-literature
| 2417665 | ecrin-mdr-crc
| S-EPMC6451045 | biostudies-literature
| S-EPMC8376946 | biostudies-literature
| S-EPMC3339890 | biostudies-other