Repeatability and Comparability of Retinal Blood Vessel Caliber Measurements by OCTA.


ABSTRACT:

Background

To investigate the repeatability in vessel caliber measurements by optical coherence tomography angiography (OCTA).

Methods

In this prospective study, 28 patients (47 eyes) underwent sequential OCTA imaging of the optic nerve head and macula. Two independent masked graders measured vessel caliber for sequential images of the optic nerve head and macula. The average vessel width was determined and variability between graders and images.

Results

A total of 8400 measurements of 420 vessels from 84 OCTA images were included in the analysis. Overall, inter-grader agreement was excellent (ICC 0.90). The coefficient of variation (CoV) for all repeated OCTA images was 0.10. Greater glaucoma severity, older age, macular location, and diagnosis of diabetes were associated with thinner vessels (p < 0.05). CoV was higher in the peripapillary region (0.07) as compared to the macula (0.15). ICC was high for all subgroups except for the macula (ICC = 0.72).

Conclusions

Overall, the repeatability of vessel caliber measurements by OCTA was high and variability low. There was greater variability in the measurement of macular vessels, possibly due to technical limitations in acquiring accurate vessel widths for smaller macular vessels.

SUBMITTER: Tsai J 

PROVIDER: S-EPMC10366731 | biostudies-literature | 2023 Jul

REPOSITORIES: biostudies-literature

altmetric image

Publications

Repeatability and Comparability of Retinal Blood Vessel Caliber Measurements by OCTA.

Tsai Joby J   Asanad Samuel S   Whiting Martha M   Zhang Xuemin X   Magder Laurence L   Saeedi Osamah O  

Vision (Basel, Switzerland) 20230703 3


<h4>Background</h4>To investigate the repeatability in vessel caliber measurements by optical coherence tomography angiography (OCTA).<h4>Methods</h4>In this prospective study, 28 patients (47 eyes) underwent sequential OCTA imaging of the optic nerve head and macula. Two independent masked graders measured vessel caliber for sequential images of the optic nerve head and macula. The average vessel width was determined and variability between graders and images.<h4>Results</h4>A total of 8400 mea  ...[more]

Similar Datasets

| S-EPMC5905988 | biostudies-literature
| S-EPMC6040728 | biostudies-literature
| S-EPMC9357695 | biostudies-literature
| S-EPMC8675673 | biostudies-literature
2009-05-14 | GSE13860 | GEO
| S-EPMC6297956 | biostudies-literature
| S-EPMC5063055 | biostudies-literature
| S-EPMC3713191 | biostudies-literature
| S-EPMC10876697 | biostudies-literature
| S-EPMC4859428 | biostudies-literature