Comparison of Common Retinal Vessel Caliber Measurement Software and a Conversion Algorithm.


ABSTRACT:

Purpose

To compare three commonly used retinal vessel caliber measurement software systems, and propose an algorithm for conversion between measurement systems.

Methods

We used 120 retinal photographs to evaluate the agreement between three commonly used software (Retinal Analysis [RA], Integrative Vessel Analysis [IVAN], and Singapore I Vessel Assessment [SIVA]). Bland-Altman plots were used to evaluate agreement of retinal arteriolar (central retinal artery equivalent, CRAE) and venular (central retinal vein equivalent, CRVE) calibers. Pearson's correlation was used to assess the associations between systemic factors and retinal vessel calibers, and Z-test was used to compare the strength of the correlation coefficients across the three software systems. An algorithm was created to convert measurements, with paired t-test performed to evaluate the differences between SIVA-measured retinal calibers and SIVA-approximates converted from RA- and IVAN-measurements using the algorithm.

Results

Differences between SIVA- and RA-measured calibers (CRAE: mean difference [MD] = -21.8 ?m, 95% limits of agreement [LOA], -47.3 to 3.7 ?m; CRVE: MD = -7.7 ?m, 95% LOA, -28.0 to 12.6 ?m), SIVA- and IVAN-measured calibers (CRAE: MD = -6.7 ?m, 95% LOA, -23.8 to 10.4 ?m; CRVE: MD = -18.2 ?m 95% LOA, -36.7 to 0.4 ?m) were large. However, the strength of correlations between systemic factors with SIVA-measured retinal calibers was not significantly different to that measured using RA and IVAN (P ? 0.332). SIVA-approximates converted from RA and IVAN measurements using the proposed algorithm was not significantly different from SIVA-measured calibers (P ? 0.20).

Conclusion

Absolute measurements of retinal vessel calibers vary between three common software systems but associations with systemic factors were similar.

Translational relevance

The proposed algorithm allowed conversions of RA and IVAN measurements to SIVA-approximates. This conversion is important for future data pooling and establishment of normative values for retinal vascular caliber measurements.

SUBMITTER: Yip W 

PROVIDER: S-EPMC5063055 | biostudies-literature | 2016 Oct

REPOSITORIES: biostudies-literature

altmetric image

Publications

Comparison of Common Retinal Vessel Caliber Measurement Software and a Conversion Algorithm.

Yip WanFen W   Tham Yih Chung YC   Hsu Wynne W   Lee Mong Li ML   Klein Ronald R   Klein Barbara B   Ikram Mohammad Kamran MK   Wong Tien Yin TY   Cheung Carol Yim-Lui CY  

Translational vision science & technology 20161012 5


<h4>Purpose</h4>To compare three commonly used retinal vessel caliber measurement software systems, and propose an algorithm for conversion between measurement systems.<h4>Methods</h4>We used 120 retinal photographs to evaluate the agreement between three commonly used software (Retinal Analysis [RA], Integrative Vessel Analysis [IVAN], and Singapore I Vessel Assessment [SIVA]). Bland-Altman plots were used to evaluate agreement of retinal arteriolar (central retinal artery equivalent, CRAE) and  ...[more]

Similar Datasets

| S-EPMC9357695 | biostudies-literature
| S-EPMC6040728 | biostudies-literature
| S-EPMC10366731 | biostudies-literature
| S-EPMC3713191 | biostudies-literature
| S-EPMC4859428 | biostudies-literature
| S-EPMC4290848 | biostudies-literature
| S-EPMC5905988 | biostudies-literature
| S-EPMC3289046 | biostudies-literature
| S-EPMC4709546 | biostudies-literature
| S-EPMC7042908 | biostudies-literature