Ontology highlight
ABSTRACT: Background
Esophageal thermal injury is a complication of atrial fibrillation (AF) ablation, and it can be avoided by esophageal deviation during left atrial posterior wall radiofrequency catheter ablation.Objective
This study aimed to evaluate the safety of a nitinol-based mechanical esophageal displacement device (MEDD) and its performance.Methods
This preclinical safety study was conducted on 20 pigs, with 10 undergoing radiofrequency AF ablation using the MEDD and 10 serving as a control group under anticoagulation but without radiofrequency application. Esophageal traumatic injuries were classified from 0 to 4 and were grouped as absent (grade 0), minor (grade 1 or 2), moderate (grade 3), or major risk lesions (grade 4) by anatomopathological study. Grades 1 and 2 were considered acceptable. Fluoroscopy was used to measure displacement.Results
Five (25%) pigs developed traumatic lesions, 4 with grade 1 and 1 with grade 2 (2-mm superficial ulcer). There was no difference in lesion occurrence between the radiofrequency and control groups (30% and 20%, respectively; P = .43). Under rightward displacement, the right edge moved 23.9 (interquartile range [IQR] 21.3-26.3) mm and the left edge moved 16.3 (IQR 13.8-18.4) mm (P < .001) from baseline. Under leftward displacement, the right edge moved 13.5 (IQR 10.9-15.3) mm and the left edge moved 16.5 (IQR 12.3-18.5) mm (P = .07). A perforation to the pharyngeal diverticulum occurred in 1 pig, related to an accidental extubation.Conclusion
In pigs, the MEDD demonstrated safety in relation to esophageal tissue, and successful deviation. Esophageal traumatic injuries were acceptable, but improper manipulation led to pharyngeal lesion.
SUBMITTER: Pereira R
PROVIDER: S-EPMC10513920 | biostudies-literature | 2023 Sep
REPOSITORIES: biostudies-literature
Pereira Renner R Pisani Cristiano C Aiello Vera V Cestari Idágene I Oyama Helena H Santos Osmar O Otubo Jorge J Moura Daniel D Scanavacca Mauricio M
Heart rhythm O2 20230809 9
<h4>Background</h4>Esophageal thermal injury is a complication of atrial fibrillation (AF) ablation, and it can be avoided by esophageal deviation during left atrial posterior wall radiofrequency catheter ablation.<h4>Objective</h4>This study aimed to evaluate the safety of a nitinol-based mechanical esophageal displacement device (MEDD) and its performance.<h4>Methods</h4>This preclinical safety study was conducted on 20 pigs, with 10 undergoing radiofrequency AF ablation using the MEDD and 10 ...[more]