Unknown

Dataset Information

0

Development of a mast cell-based biosensor.


ABSTRACT: A mast cell-based biosensor has been developed to enable the use of these cells in numerous applications including pharmaceutical screening, environmental monitoring, clinical diagnosis and homeland security. Rat basophilic leukemia (RBL) mast cells offer excellent potential for biosensor applications because they are robust and undergo a dramatic exocytotic response within minutes of antigen addition. To monitor mast cell activation, fluorescent dyes were loaded into the cells and used as indicators of alkalinization of secretory granules, calcium fluxes or generation of reactive oxygen species. These fluorescence assays efficiently measure activation of antigen-stimulated RBL mast cells, detecting the antigen with picomolar sensitivity. To demonstrate the utility of this mast cell-based biosensor for detection of microbial pathogens, an IgE chimeric protein was created by fusing the Fc region of the IgE antibody to CD14, a receptor for lipopolysaccharide. This chimeric protein has the capacity to bind to Escherichia coli and Listeria monocytogenes and also to IgE receptors on the mast cells, thereby stimulating a signaling response to bacteria. RBL mast cells labeled with the calcium indicator Fluo-4 are shown to be responsive to E. coli, only when sensitized with the chimeric protein, thus demonstrating a highly versatile biosensor for bacterial contamination.

SUBMITTER: Curtis T 

PROVIDER: S-EPMC2269709 | biostudies-literature | 2008 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Development of a mast cell-based biosensor.

Curtis Theresa T   Naal Rose Mary Z G RM   Batt Carl C   Tabb Joel J   Holowka David D  

Biosensors & bioelectronics 20071024 7


A mast cell-based biosensor has been developed to enable the use of these cells in numerous applications including pharmaceutical screening, environmental monitoring, clinical diagnosis and homeland security. Rat basophilic leukemia (RBL) mast cells offer excellent potential for biosensor applications because they are robust and undergo a dramatic exocytotic response within minutes of antigen addition. To monitor mast cell activation, fluorescent dyes were loaded into the cells and used as indic  ...[more]

Similar Datasets

| S-EPMC5778388 | biostudies-literature
| S-EPMC7825672 | biostudies-literature
| S-EPMC10294642 | biostudies-literature
| S-EPMC2891049 | biostudies-literature
| S-EPMC7606717 | biostudies-literature
| S-EPMC5663365 | biostudies-literature
| S-EPMC5659818 | biostudies-literature
| S-EPMC4219169 | biostudies-other
| S-EPMC4955689 | biostudies-literature
| S-EPMC7948755 | biostudies-literature