Theory and validation of magnetic resonance fluid motion estimation using intensity flow data.


ABSTRACT:

Background

Motion tracking based on spatial-temporal radio-frequency signals from the pixel representation of magnetic resonance (MR) imaging of a non-stationary fluid is able to provide two dimensional vector field maps. This supports the underlying fundamentals of magnetic resonance fluid motion estimation and generates a new methodology for flow measurement that is based on registration of nuclear signals from moving hydrogen nuclei in fluid. However, there is a need to validate the computational aspect of the approach by using velocity flow field data that we will assume as the true reference information or ground truth.

Methodology/principal findings

In this study, we create flow vectors based on an ideal analytical vortex, and generate artificial signal-motion image data to verify our computational approach. The analytical and computed flow fields are compared to provide an error estimate of our methodology. The comparison shows that the fluid motion estimation approach using simulated MR data is accurate and robust enough for flow field mapping. To verify our methodology, we have tested the computational configuration on magnetic resonance images of cardiac blood and proved that the theory of magnetic resonance fluid motion estimation can be applicable practically.

Conclusions/significance

The results of this work will allow us to progress further in the investigation of fluid motion prediction based on imaging modalities that do not require velocity encoding. This article describes a novel theory of motion estimation based on magnetic resonating blood, which may be directly applied to cardiac flow imaging.

SUBMITTER: Wong KK 

PROVIDER: S-EPMC2651647 | biostudies-literature | 2009

REPOSITORIES: biostudies-literature

altmetric image

Publications

Theory and validation of magnetic resonance fluid motion estimation using intensity flow data.

Wong Kelvin Kian Loong KK   Kelso Richard Malcolm RM   Worthley Stephen Grant SG   Sanders Prashanthan P   Mazumdar Jagannath J   Abbott Derek D  

PloS one 20090309 3


<h4>Background</h4>Motion tracking based on spatial-temporal radio-frequency signals from the pixel representation of magnetic resonance (MR) imaging of a non-stationary fluid is able to provide two dimensional vector field maps. This supports the underlying fundamentals of magnetic resonance fluid motion estimation and generates a new methodology for flow measurement that is based on registration of nuclear signals from moving hydrogen nuclei in fluid. However, there is a need to validate the c  ...[more]

Similar Datasets

| S-EPMC2682657 | biostudies-literature
| S-EPMC7477913 | biostudies-literature
| S-EPMC7324609 | biostudies-literature
| S-EPMC3146240 | biostudies-literature
| S-EPMC6519164 | biostudies-literature
| S-EPMC7669675 | biostudies-literature
| S-EPMC6170141 | biostudies-literature
| S-EPMC6802342 | biostudies-literature
| S-EPMC3895106 | biostudies-literature