ABSTRACT: Aims
We investigated the feasibility of real-time magnetic resonance imaging (RTMRI) guided ablation of the cavotricuspid isthmus (CTI) by using a MRI-compatible ablation catheter.Methods and results
Cavotricuspid isthmus ablation was performed in an interventional RTMRI suite by using a novel 7 French, steerable, non-ferromagnetic ablation catheter in a porcine in vivo model (n = 20). The catheter was introduced and navigated by RTMRI visualization only. Catheter position and movement during manipulation were continuously visualized during the entire intervention. Two porcine prematurely died due to VT/VF. Anatomical completion of the CTI ablation line could be achieved after a mean of 6.3+/-3 RF pulses (RF energy: 1807+/-1016.4 Ws/RF pulse, temperature: 55.9+/-5.9 degrees C) in n = 18 animals. In 15 of 18 procedures (83.3%) a complete CTI block was proven by conventional mapping in the electrophysiological (EP) lab.Conclusion
Completely non-fluoroscopic ablation guided by RTMRI using a steerable and non-ferromagnetic catheter is a promising novel technology in interventional electrophysiology.
SUBMITTER: Hoffmann BA
PROVIDER: S-EPMC2821629 | biostudies-literature | 2010 Feb
REPOSITORIES: biostudies-literature
Hoffmann Boris A BA Koops Andreas A Rostock Thomas T Müllerleile Kai K Steven Daniel D Karst Roman R Steinke Mark U MU Drewitz Imke I Lund Gunnar G Koops Susan S Adam Gerhard G Willems Stephan S
European heart journal 20091105 4
<h4>Aims</h4>We investigated the feasibility of real-time magnetic resonance imaging (RTMRI) guided ablation of the cavotricuspid isthmus (CTI) by using a MRI-compatible ablation catheter.<h4>Methods and results</h4>Cavotricuspid isthmus ablation was performed in an interventional RTMRI suite by using a novel 7 French, steerable, non-ferromagnetic ablation catheter in a porcine in vivo model (n = 20). The catheter was introduced and navigated by RTMRI visualization only. Catheter position and mo ...[more]