Partial volume correction incorporating Rb-82 positron range for quantitative myocardial perfusion PET based on systolic-diastolic activity ratios and phantom measurements.


ABSTRACT:

Background

Quantitative myocardial PET perfusion imaging requires partial volume corrections.

Methods

Patients underwent ECG-gated, rest-dipyridamole, myocardial perfusion PET using Rb-82 decay corrected in Bq/cc for diastolic, systolic, and combined whole cycle ungated images. Diastolic partial volume correction relative to systole was determined from the systolic/diastolic activity ratio, systolic partial volume correction from phantom dimensions comparable to systolic LV wall thicknesses and whole heart cycle partial volume correction for ungated images from fractional systolic-diastolic duration for systolic and diastolic partial volume corrections.

Results

For 264 PET perfusion images from 159 patients (105 rest-stress image pairs, 54 individual rest or stress images), average resting diastolic partial volume correction relative to systole was 1.14 ± 0.04, independent of heart rate and within ±1.8% of stress images (1.16 ± 0.04). Diastolic partial volume corrections combined with those for phantom dimensions comparable to systolic LV wall thickness gave an average whole heart cycle partial volume correction for ungated images of 1.23 for Rb-82 compared to 1.14 if positron range were negligible as for F-18.

Conclusion

Quantitative myocardial PET perfusion imaging requires partial volume correction, herein demonstrated clinically from systolic/diastolic absolute activity ratios combined with phantom data accounting for Rb-82 positron range.

SUBMITTER: Johnson NP 

PROVIDER: S-EPMC3069317 | biostudies-literature | 2011 Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

Partial volume correction incorporating Rb-82 positron range for quantitative myocardial perfusion PET based on systolic-diastolic activity ratios and phantom measurements.

Johnson Nils P NP   Sdringola Stefano S   Gould K Lance KL  

Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology 20101224 2


<h4>Background</h4>Quantitative myocardial PET perfusion imaging requires partial volume corrections.<h4>Methods</h4>Patients underwent ECG-gated, rest-dipyridamole, myocardial perfusion PET using Rb-82 decay corrected in Bq/cc for diastolic, systolic, and combined whole cycle ungated images. Diastolic partial volume correction relative to systole was determined from the systolic/diastolic activity ratio, systolic partial volume correction from phantom dimensions comparable to systolic LV wall t  ...[more]

Similar Datasets

| S-EPMC6431039 | biostudies-literature
| S-EPMC3223511 | biostudies-other
| S-EPMC4859355 | biostudies-literature
| S-EPMC10749049 | biostudies-literature
| S-EPMC6430746 | biostudies-literature
| S-EPMC5556151 | biostudies-other
| S-EPMC8873116 | biostudies-literature
| S-EPMC5628109 | biostudies-literature
| S-EPMC8581512 | biostudies-literature
| S-EPMC5646228 | biostudies-literature