Unknown

Dataset Information

0

Unusual interferon gamma measurements with QuantiFERON-TB Gold and QuantiFERON-TB Gold In-Tube tests.


ABSTRACT:

Introduction

Interferon gamma (IFN-?) release assays, such as QuantiFERON®-TB Gold test (QFT-G) and QuantiFERON®-TB Gold In-Tube test (QFT-GIT) are designed to detect M. tuberculosis (Mtb) infection. Recognition of unusual IFN-? measurements may help indicate inaccurate results.

Methods

We examined QFT-G and QFT-GIT results from subjects who had two or more tests completed. We classified unusual IFN-? measurements as: 1) High Nil Concentration (HNC) when IFN-? concentration in plasma from unstimulated blood exceeded 0.7 IU/mL; 2) Low Mitogen Response (LMR) when Mitogen Response was <0.5 IU/mL; 3) Very Low Mitogen Response (VLMR) when Mitogen Response was ?-0.5 IU/mL; and 4) Very Low Antigen Response (VLAR) when the response to a Mtb antigen was ?-0.35 IU/mL and ?-0.5 times the IFN-? concentration in plasma from unstimulated blood.

Results

Among 5,309 results from 1,728 subjects, HNC occurred in 234 (4.4%) tests for 162 subjects, LMR in 108 (2.0%) tests for 85 subjects, VLMR in 22 (0.4%) tests for 21 subjects, and VLAR in 41 (0.8%) tests for 39 subjects. QFT-GIT had fewer HNC, VLMR, and VLAR (p?=?0.042, 0.004, and 0.067 respectively); QFT-G had fewer LMR (p?=?0.005). Twenty-four (51.6%) of 47 subjects with positive results and HNC were negative or indeterminate by all other tests. Thirteen (61.9%) of 21 subjects with positive results and LMR were negative or indeterminate by all other tests.

Conclusion

Unusual IFN-? measurements including HNC, LMR, VLMR, and VLAR were encountered in small numbers, and in most instances were not seen on simultaneously or subsequently performed tests. To avoid erroneous diagnosis of Mtb infection, IGRAs with unusual IFN-? measurements should be repeated with another blood sample and interpreted with caution if they recur.

SUBMITTER: Powell RD 

PROVIDER: S-EPMC3110578 | biostudies-literature | 2011

REPOSITORIES: biostudies-literature

altmetric image

Publications

Unusual interferon gamma measurements with QuantiFERON-TB Gold and QuantiFERON-TB Gold In-Tube tests.

Powell Richard D RD   Whitworth William C WC   Bernardo John J   Moonan Patrick K PK   Mazurek Gerald H GH  

PloS one 20110608 6


<h4>Introduction</h4>Interferon gamma (IFN-γ) release assays, such as QuantiFERON®-TB Gold test (QFT-G) and QuantiFERON®-TB Gold In-Tube test (QFT-GIT) are designed to detect M. tuberculosis (Mtb) infection. Recognition of unusual IFN-γ measurements may help indicate inaccurate results.<h4>Methods</h4>We examined QFT-G and QFT-GIT results from subjects who had two or more tests completed. We classified unusual IFN-γ measurements as: 1) High Nil Concentration (HNC) when IFN-γ concentration in pla  ...[more]

Similar Datasets

| S-EPMC3435391 | biostudies-literature
| S-EPMC3570654 | biostudies-literature
| S-EPMC6813010 | biostudies-literature
| S-EPMC5697869 | biostudies-literature
| S-EPMC3900587 | biostudies-literature