Agreement analysis of LENSTAR with other techniques of biometry.


ABSTRACT: PURPOSE: To assess the agreement of the optical low-coherence reflectometry (OLCR) device LENSTAR LS900 with partial coherence interferometry (PCI) device IOLMaster and applanation and immersion ultrasound biometry. METHODS: We conducted the study at the Ophthalmology Clinic, University of Malaya Medical Center, Malaysia. Phakic eyes of 76 consecutive cataract patients were measured using four different methods: IOLMaster, LENSTAR and A-scan applanation and immersion ultrasound biometry. We assessed the method agreement in the LENSTAR-IOLMaster, LENSTAR-applanation, and LENSTAR-immersion comparisons for axial length (AL) and intraocular lens (IOL) power using Bland-Altman plots. For average K, we compared LENSTAR with IOLMaster and the TOPCON KR-8100 autorefractor-keratometer. SRK/T formula was used to compute IOL power, with emmetropia as the target refractive outcome. RESULTS: For all the variables studied, LENSTAR agreement with IOLMaster is strongest, followed by those with immersion and applanation. For the LENSTAR-IOLMaster comparison, the estimated proportion of differences falling within 0.33?mm from zero AL and within 1D from zero IOL power is 100%. The estimated proportion of differences falling within 0.5 D from zero average K is almost 100% in the LENSTAR-IOLMaster comparison but 88% in the LENSTAR-TOPCON comparison. The proportion of differences falling within 0.10?mm (AL) and within 1D (IOL power) in the LENSTAR-IOLMaster comparison has practically significant discrepancy with that of LENSTAR-applanation and LENSTAR-immersion comparisons. CONCLUSIONS: In phakic eyes of cataract patients, measurements of AL, average K, and IOL power calculated using the SRK/T formula from LENSTAR are biometrically equivalent to those from IOLMaster, but not with those from applanation and immersion ultrasound biometry.

SUBMITTER: Jasvinder S 

PROVIDER: S-EPMC3178124 | biostudies-literature | 2011 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Agreement analysis of LENSTAR with other techniques of biometry.

Jasvinder S S   Khang T F TF   Sarinder K K S KK   Loo V P VP   Subrayan V V  

Eye (London, England) 20110311 6


<h4>Purpose</h4>To assess the agreement of the optical low-coherence reflectometry (OLCR) device LENSTAR LS900 with partial coherence interferometry (PCI) device IOLMaster and applanation and immersion ultrasound biometry.<h4>Methods</h4>We conducted the study at the Ophthalmology Clinic, University of Malaya Medical Center, Malaysia. Phakic eyes of 76 consecutive cataract patients were measured using four different methods: IOLMaster, LENSTAR and A-scan applanation and immersion ultrasound biom  ...[more]

Similar Datasets

| S-EPMC5546377 | biostudies-other
| S-EPMC8720848 | biostudies-literature
| S-EPMC8284630 | biostudies-literature
| S-EPMC5504265 | biostudies-literature
| S-EPMC4304259 | biostudies-literature
| S-EPMC3637028 | biostudies-literature
| S-EPMC6048584 | biostudies-literature
| S-EPMC5021199 | biostudies-literature
| S-EPMC6219795 | biostudies-literature
| S-EPMC4895813 | biostudies-literature