Unknown

Dataset Information

0

Accuracy of CT-based attenuation correction in PET/CT bone imaging.


ABSTRACT: We evaluate the accuracy of scaling CT images for attenuation correction of PET data measured for bone. While the standard tri-linear approach has been well tested for soft tissues, the impact of CT-based attenuation correction on the accuracy of tracer uptake in bone has not been reported in detail. We measured the accuracy of attenuation coefficients of bovine femur segments and patient data using a tri-linear method applied to CT images obtained at different kVp settings. Attenuation values at 511 keV obtained with a (68)Ga/(68)Ge transmission scan were used as a reference standard. The impact of inaccurate attenuation images on PET standardized uptake values (SUVs) was then evaluated using simulated emission images and emission images from five patients with elevated levels of FDG uptake in bone at disease sites. The CT-based linear attenuation images of the bovine femur segments underestimated the true values by 2.9 ± 0.3% for cancellous bone regardless of kVp. For compact bone the underestimation ranged from 1.3% at 140 kVp to 14.1% at 80 kVp. In the patient scans at 140 kVp the underestimation was approximately 2% averaged over all bony regions. The sensitivity analysis indicated that errors in PET SUVs in bone are approximately proportional to errors in the estimated attenuation coefficients for the same regions. The variability in SUV bias also increased approximately linearly with the error in linear attenuation coefficients. These results suggest that bias in bone uptake SUVs of PET tracers ranges from 2.4% to 5.9% when using CT scans at 140 and 120 kVp for attenuation correction. Lower kVp scans have the potential for considerably more error in dense bone. This bias is present in any PET tracer with bone uptake but may be clinically insignificant for many imaging tasks. However, errors from CT-based attenuation correction methods should be carefully evaluated if quantitation of tracer uptake in bone is important.

SUBMITTER: Abella M 

PROVIDER: S-EPMC3353415 | biostudies-literature | 2012 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

Accuracy of CT-based attenuation correction in PET/CT bone imaging.

Abella Monica M   Alessio Adam M AM   Mankoff David A DA   MacDonald Lawrence R LR   Vaquero Juan Jose JJ   Desco Manuel M   Kinahan Paul E PE  

Physics in medicine and biology 20120405 9


We evaluate the accuracy of scaling CT images for attenuation correction of PET data measured for bone. While the standard tri-linear approach has been well tested for soft tissues, the impact of CT-based attenuation correction on the accuracy of tracer uptake in bone has not been reported in detail. We measured the accuracy of attenuation coefficients of bovine femur segments and patient data using a tri-linear method applied to CT images obtained at different kVp settings. Attenuation values a  ...[more]

Similar Datasets

| S-EPMC4090355 | biostudies-other
| S-EPMC10702200 | biostudies-literature
| S-EPMC3869851 | biostudies-other
| S-EPMC3635797 | biostudies-other
| S-EPMC6779234 | biostudies-literature
| S-EPMC6660490 | biostudies-literature
| S-EPMC5628109 | biostudies-literature
| S-EPMC9537165 | biostudies-literature
| S-EPMC6910626 | biostudies-literature
| S-EPMC4891225 | biostudies-literature