Unknown

Dataset Information

0

Impedance and electrically evoked compound action potential (ECAP) drop within 24 hours after cochlear implantation.


ABSTRACT: Previous animal study revealed that post-implantation electrical detection levels significantly declined within days. The impact of cochlear implant (CI) insertion on human auditory pathway in terms of impedance and electrically evoked compound action potential (ECAP) variation within hours after surgery remains unclear, since at this time frequency mapping can only commence weeks after implantation due to factors associated with wound conditions. The study presented our experiences with regards to initial switch-on within 24 hours, and thus the findings about the milieus inside cochlea within the first few hours after cochlear implantation in terms of impedance/ECAP fluctuations. The charts of fifty-four subjects with profound hearing impairment were studied. A minimal invasive approach was used for cochlear implantation, characterized by a small skin incision (? 2.5 cm) and soft techniques for cochleostomy. Impedance/ECAP was measured intro-operatively and within 24 hours post-operatively. Initial mapping within 24 hours post-operatively was performed in all patients without major complications. Impedance/ECAP became significantly lower measured within 24 hours post-operatively as compared with intra-operatively (p<0.001). There were no differences between pre-operative and post-operative threshold for air-conduction hearing. A significant drop of impedance/ECAP in one day after cochlear implantation was revealed for the first time in human beings. Mechanisms could be related to the restoration of neuronal sensitivity to the electrical stimulation, and/or the interaction between the matrix enveloping the electrodes and the electrical stimulation of the initial switch-on. Less wound pain/swelling and soft techniques both contributed to the success of immediate initial mapping, which implied a stable micro-environment inside the cochlea despite electrodes insertion. Our research invites further studies to correlate initial impedance/ECAP changes with long-term hearing/speech performance.

SUBMITTER: Chen JK 

PROVIDER: S-EPMC3753283 | biostudies-literature | 2013

REPOSITORIES: biostudies-literature

altmetric image

Publications

Impedance and electrically evoked compound action potential (ECAP) drop within 24 hours after cochlear implantation.

Chen Joshua Kuang-Chao JK   Chuang Ann Yi-Chiun AY   Sprinzl Georg Mathias GM   Tung Tao-Hsin TH   Li Lieber Po-Hung LP  

PloS one 20130826 8


Previous animal study revealed that post-implantation electrical detection levels significantly declined within days. The impact of cochlear implant (CI) insertion on human auditory pathway in terms of impedance and electrically evoked compound action potential (ECAP) variation within hours after surgery remains unclear, since at this time frequency mapping can only commence weeks after implantation due to factors associated with wound conditions. The study presented our experiences with regards  ...[more]

Similar Datasets

| S-EPMC6752796 | biostudies-literature
| S-EPMC8611070 | biostudies-literature
| S-EPMC7026160 | biostudies-literature
| S-EPMC5481377 | biostudies-other
| S-EPMC3709093 | biostudies-literature
| S-EPMC7198904 | biostudies-literature
| S-EPMC5404966 | biostudies-literature
| S-EPMC7338891 | biostudies-literature
| S-EPMC8263475 | biostudies-literature
| S-EPMC5748379 | biostudies-literature