Variability of the QuantiFERON®-TB gold in-tube test using automated and manual methods.


ABSTRACT:

Background

The QuantiFERON®-TB Gold In-Tube test (QFT-GIT) detects Mycobacterium tuberculosis (Mtb) infection by measuring release of interferon gamma (IFN-?) when T-cells (in heparinized whole blood) are stimulated with specific Mtb antigens. The amount of IFN-? is determined by enzyme-linked immunosorbent assay (ELISA). Automation of the ELISA method may reduce variability. To assess the impact of ELISA automation, we compared QFT-GIT results and variability when ELISAs were performed manually and with automation.

Methods

Blood was collected into two sets of QFT-GIT tubes and processed at the same time. For each set, IFN-? was measured in automated and manual ELISAs. Variability in interpretations and IFN-? measurements was assessed between automated (A1 vs. A2) and manual (M1 vs. M2) ELISAs. Variability in IFN-? measurements was also assessed on separate groups stratified by the mean of the four ELISAs.

Results

Subjects (N?=?146) had two automated and two manual ELISAs completed. Overall, interpretations were discordant for 16 (11%) subjects. Excluding one subject with indeterminate results, 7 (4.8%) subjects had discordant automated interpretations and 10 (6.9%) subjects had discordant manual interpretations (p?=?0.17). Quantitative variability was not uniform; within-subject variability was greater with higher IFN-? measurements and with manual ELISAs. For subjects with mean TB Responses ±0.25 IU/mL of the 0.35 IU/mL cutoff, the within-subject standard deviation for two manual tests was 0.27 (CI95?=?0.22-0.37) IU/mL vs. 0.09 (CI95?=?0.07-0.12) IU/mL for two automated tests.

Conclusion

QFT-GIT ELISA automation may reduce variability near the test cutoff. Methodological differences should be considered when interpreting and using IFN-? release assays (IGRAs).

SUBMITTER: Whitworth WC 

PROVIDER: S-EPMC3900587 | biostudies-literature | 2014

REPOSITORIES: biostudies-literature

altmetric image

Publications


<h4>Background</h4>The QuantiFERON®-TB Gold In-Tube test (QFT-GIT) detects Mycobacterium tuberculosis (Mtb) infection by measuring release of interferon gamma (IFN-γ) when T-cells (in heparinized whole blood) are stimulated with specific Mtb antigens. The amount of IFN-γ is determined by enzyme-linked immunosorbent assay (ELISA). Automation of the ELISA method may reduce variability. To assess the impact of ELISA automation, we compared QFT-GIT results and variability when ELISAs were performed  ...[more]

Similar Datasets

| S-EPMC3570654 | biostudies-literature
| S-EPMC3435391 | biostudies-literature
| S-EPMC3110578 | biostudies-literature
| S-EPMC5697869 | biostudies-literature
| S-EPMC4227090 | biostudies-other
| S-EPMC6204680 | biostudies-literature
| S-EPMC7297367 | biostudies-literature
| S-EPMC3395691 | biostudies-literature
| S-EPMC3370442 | biostudies-other
| S-EPMC6070176 | biostudies-literature