ABSTRACT: Purpose
Determining the magnitude of displacement in pediatric lateral humeral condyle fractures can be difficult. The purpose of this study was to (1) assess the effect of forearm rotation on true fracture displacement using a cadaver model and to (2) determine the accuracy of radiographic measurements of the fracture gap.Methods
A non-displaced fracture was created in three human cadaveric arms. The specimens were mounted on a custom apparatus allowing forearm rotation with the humerus fixed. First, the effect of pure rotation on fracture displacement was simulated by rotating the forearm from supination to pronation about the central axis of the forearm, to isolate the effects of muscle pull. Then, the clinical condition of obtaining a lateral oblique radiograph was simulated by rotating the forearm about the medial aspect of the forearm. Fracture displacements were measured using a motion-capture system (true-displacement) and clinical radiographs (apparent-displacement).Results
During pure rotation of the forearm, there were no significant differences in fracture displacement between supination and pronation, with changes in displacement of <1.0 mm. During rotation about the medial aspect of the forearm, there was a significant difference in true displacements between supination and pronation at the posterior edge (p < 0.05).Conclusion
Overall, true fracture displacement measurements were larger than apparent radiographic displacement measurements, with differences from 1.6 to 6.0 mm, suggesting that the current clinical methods may not be sensitive enough to detect a displacement of 2.0 mm, especially when positioning the upper extremity for an internal oblique lateral radiograph.
SUBMITTER: Knutsen A
PROVIDER: S-EPMC3935020 | biostudies-literature |
REPOSITORIES: biostudies-literature