A new microfluidics-based droplet dispenser for ICPMS.


ABSTRACT: In this work, a novel droplet microfluidic sample introduction system for inductively coupled plasma mass spectrometry (ICPMS) is proposed and characterized. The cheap and disposable microfluidic chip generates droplets of an aqueous sample in a stream of perfluorohexane (PFH), which is also used to eject them as a liquid jet. The aqueous droplets remain intact during the ejection and can be transported into the ICP with >50% efficiency. The transport is realized via a custom-built system, which includes a membrane desolvator necessary for the PFH vapor removal. The introduction system presented here can generate highly monodisperse droplets in the size range of 40-60 ?m at frequencies from 90 to 300 Hz. These droplets produced very stable signals with a relative standard deviation (RSD) comparable to the one achieved with a commercial droplet dispenser. Using the current system, samples with a total volume of <1 ?L can be analyzed. Moreover, the capabilities of the setup for introduction and quantitative elemental analysis of single cells were described using a test system of bovine red blood cells. In the future, other modules of the modern microfludics can be integrated in the chip, such as on-chip sample pretreatment or parallel introduction of different samples.

SUBMITTER: Verboket PE 

PROVIDER: S-EPMC4063494 | biostudies-literature | 2014 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

A new microfluidics-based droplet dispenser for ICPMS.

Verboket Pascal E PE   Borovinskaya Olga O   Meyer Nicole N   Günther Detlef D   Dittrich Petra S PS  

Analytical chemistry 20140531 12


In this work, a novel droplet microfluidic sample introduction system for inductively coupled plasma mass spectrometry (ICPMS) is proposed and characterized. The cheap and disposable microfluidic chip generates droplets of an aqueous sample in a stream of perfluorohexane (PFH), which is also used to eject them as a liquid jet. The aqueous droplets remain intact during the ejection and can be transported into the ICP with >50% efficiency. The transport is realized via a custom-built system, which  ...[more]

Similar Datasets

| S-EPMC8748332 | biostudies-literature
| S-EPMC3690483 | biostudies-literature
| S-EPMC7283233 | biostudies-literature
| S-EPMC7531118 | biostudies-literature
| S-EPMC3360709 | biostudies-other
| S-EPMC9400878 | biostudies-literature
| S-EPMC4128248 | biostudies-literature
| S-EPMC3258639 | biostudies-literature
| S-EPMC5431057 | biostudies-literature
| S-EPMC3406880 | biostudies-other