Unknown

Dataset Information

0

The effects of computed tomography image characteristics and knot spacing on the spatial accuracy of B-spline deformable image registration in the head and neck geometry.


ABSTRACT: OBJECTIVES: To explore the effects of computed tomography (CT) image characteristics and B-spline knot spacing (BKS) on the spatial accuracy of a B-spline deformable image registration (DIR) in the head-and-neck geometry. METHODS: The effect of image feature content, image contrast, noise, and BKS on the spatial accuracy of a B-spline DIR was studied. Phantom images were created with varying feature content and varying contrast-to-noise ratio (CNR), and deformed using a known smooth B-spline deformation. Subsequently, the deformed images were repeatedly registered with the original images using different BKSs. The quality of the DIR was expressed as the mean residual displacement (MRD) between the known imposed deformation and the result of the B-spline DIR.Finally, for three patients, head-and-neck planning CT scans were deformed with a realistic deformation field derived from a rescan CT of the same patient, resulting in a simulated deformed image and an a-priori known deformation field. Hence, a B-spline DIR was performed between the simulated image and the planning CT at different BKSs. Similar to the phantom cases, the DIR accuracy was evaluated by means of MRD. RESULTS: In total, 162 phantom registrations were performed with varying CNR and BKSs. MRD-values < 1.0 mm were observed with a BKS between 10-20 mm for image contrast ? ± 250 HU and noise < ± 200 HU. Decreasing the image feature content resulted in increased MRD-values at all BKSs. Using BKS = 15 mm for the three clinical cases resulted in an average MRD < 1.0 mm. CONCLUSIONS: For synthetically generated phantoms and three real CT cases the highest DIR accuracy was obtained for a BKS between 10-20 mm. The accuracy decreased with decreasing image feature content, decreasing image contrast, and higher noise levels. Our results indicate that DIR accuracy in clinical CT images (typical noise levels < ± 100 HU) will not be effected by the amount of image noise.

SUBMITTER: Brouwer CL 

PROVIDER: S-EPMC4128373 | biostudies-literature | 2014

REPOSITORIES: biostudies-literature

altmetric image

Publications

The effects of computed tomography image characteristics and knot spacing on the spatial accuracy of B-spline deformable image registration in the head and neck geometry.

Brouwer Charlotte L CL   Kierkels Roel G J RG   van 't Veld Aart A AA   Sijtsema Nanna M NM   Meertens Harm H  

Radiation oncology (London, England) 20140729


<h4>Objectives</h4>To explore the effects of computed tomography (CT) image characteristics and B-spline knot spacing (BKS) on the spatial accuracy of a B-spline deformable image registration (DIR) in the head-and-neck geometry.<h4>Methods</h4>The effect of image feature content, image contrast, noise, and BKS on the spatial accuracy of a B-spline DIR was studied. Phantom images were created with varying feature content and varying contrast-to-noise ratio (CNR), and deformed using a known smooth  ...[more]

Similar Datasets

| S-EPMC6331282 | biostudies-literature
| S-EPMC5690934 | biostudies-literature
| S-EPMC7807680 | biostudies-literature
| S-EPMC8130232 | biostudies-literature
| S-EPMC3745275 | biostudies-other
| S-EPMC9983883 | biostudies-literature
| S-EPMC6964750 | biostudies-literature
| S-EPMC10415941 | biostudies-literature
| S-EPMC3843756 | biostudies-other
| S-EPMC9269460 | biostudies-literature