Capillary zone electrophoresis-electrospray ionization-tandem mass spectrometry for quantitative parallel reaction monitoring of peptide abundance and single-shot proteomic analysis of a human cell line.
Ontology highlight
ABSTRACT: We coupled capillary zone electrophoresis (CZE) with an ultrasensitive electrokinetically pumped nanospray ionization source for tandem mass spectrometry (MS/MS) analysis of complex proteomes. We first used the system for the parallel reaction monitoring (PRM) analysis of angiotensin II spiked in 0.45mg/mL of bovine serum albumin (BSA) digest. A calibration curve was generated between the loading amount of angiotensin II and intensity of angiotensin II fragment ions. CZE-PRM generated a linear calibration curve across over 4.5 orders of magnitude dynamic range corresponding to angiotensin II loading amount from 2amole to 150fmole. The relative standard deviations (RSDs) of migration time were <4% and the RSDs of fragment ion intensity were ?20% or less except 150fmole angiotensin II loading amount data (?36% RSD). We further applied the system for the first bottom up proteomic analysis of a human cell line using CZE-MS/MS. We generated 283 protein identifications from a 1h long, single-shot CZE MS/MS analysis of the MCF7 breast cancer cell line digest, corresponding to ?80ng loading amount. The MCF7 digest was fractionated using a C18 solid phase extraction column; single-shot analysis of a single fraction resulted in 468 protein identifications, which is by far the largest number of protein identifications reported for a mammalian proteomic sample using CZE.
SUBMITTER: Sun L
PROVIDER: S-EPMC4141985 | biostudies-literature | 2014 Sep
REPOSITORIES: biostudies-literature
ACCESS DATA