CT attenuation values of blood and myocardium: rationale for accurate coronary artery calcifications detection with multi-detector CT.


ABSTRACT:

Objectives

To determine inter-session and intra/inter-individual variations of the attenuations of aortic blood/myocardium with MDCT in the context of calcium scoring. To evaluate whether these variations are dependent on patients' characteristics.

Methods

Fifty-four volunteers were evaluated with calcium scoring non-enhanced CT. We measured attenuations (inter-individual variation) and standard deviations (SD, intra-individual variation) of the blood in the ascending aorta and of the myocardium of left ventricle. Every volunteer was examined twice to study the inter-session variation. The fat pad thickness at the sternum and noise (SD of air) were measured too. These values were correlated with the measured aortic/ventricular attenuations and their SDs (Pearson). Historically fixed thresholds (90 and 130 HU) were tested against different models based on attenuations of blood/ventricle.

Results

The mean attenuation was 46 HU (range, 17-84 HU) with mean SD 23 HU for the blood, and 39 HU (10-82 HU) with mean SD 18 HU for the myocardium. The attenuation/SD of the blood were significantly higher than those of the myocardium (p < 0.01). The inter-session variation was not significant. There was a poor correlation between SD of aortic blood/ventricle with fat thickness/noise. Based on existing models, 90 HU threshold offers a confidence interval of approximately 95% and 130 HU more than 99%.

Conclusions

Historical thresholds offer high confidence intervals for exclusion of aortic blood/myocardium and by the way for detecting calcifications. Nevertheless, considering the large variations of blood/myocardium CT values and the influence of patient's characteristics, a better approach might be an adaptive threshold.

SUBMITTER: Qanadli SD 

PROVIDER: S-EPMC4397043 | biostudies-literature | 2015

REPOSITORIES: biostudies-literature

altmetric image

Publications

CT attenuation values of blood and myocardium: rationale for accurate coronary artery calcifications detection with multi-detector CT.

Qanadli Salah D SD   Jouannic Anne-Marie AM   Dehmeshki Jamshid J   Lu Tri-Linh TL  

PloS one 20150414 4


<h4>Objectives</h4>To determine inter-session and intra/inter-individual variations of the attenuations of aortic blood/myocardium with MDCT in the context of calcium scoring. To evaluate whether these variations are dependent on patients' characteristics.<h4>Methods</h4>Fifty-four volunteers were evaluated with calcium scoring non-enhanced CT. We measured attenuations (inter-individual variation) and standard deviations (SD, intra-individual variation) of the blood in the ascending aorta and of  ...[more]

Similar Datasets

| S-EPMC10964482 | biostudies-literature
| S-EPMC8043535 | biostudies-literature
| S-EPMC8593585 | biostudies-literature
| S-EPMC7599179 | biostudies-literature
| S-EPMC8506747 | biostudies-literature
| S-EPMC8589749 | biostudies-literature
| S-EPMC8514719 | biostudies-literature
| S-EPMC10703905 | biostudies-literature
| S-EPMC7977955 | biostudies-literature
| S-EPMC7656362 | biostudies-literature