ABSTRACT: Purpose
To compare the diagnostic performance of four tracer kinetic analysis methods to quantify myocardial perfusion from magnetic resonance (MR) imaging cardiac perfusion data sets in terms of their ability to lead to the diagnosis of myocardial ischemia.Materials and methods
The study was approved by the regional ethics committee, and all patients gave written consent. A representative sample of 50 patients with suspected ischemic heart disease was retrospectively selected from the Clinical Evaluation of Magnetic Resonance Imaging in Coronary Heart Disease trial data set. Quantitative myocardial blood flow (MBF) was estimated from rest and adenosine stress MR imaging perfusion data sets by using four established methods. A matching diagnosis of both an inducible defect as assessed with single photon emission computed tomography and a luminal stenosis of 70% or more as assessed with quantitative x-ray angiography was used as the reference standard for the presence of myocardial ischemia. Diagnostic performance was evaluated with receiver operating characteristic (ROC) curve analysis for each method, with stress MBF and myocardial perfusion reserve (MPR) serving as continuous measures.Results
Area under the ROC curve with stress MBF and MPR as the outcome measures, respectively, was 0.86 and 0.92 for the Fermi model, 0.85 and 0.87 for the uptake model, 0.85 and 0.80 for the one-compartment model, and 0.87 and 0.87 for model-independent deconvolution. There was no significant difference between any of the models or between MBF and MPR, except that the Fermi model outperformed the one-compartment model if MPR was used as the outcome measure (P = .02).Conclusion
Diagnostic performance of quantitative myocardial perfusion estimates is not affected by the tracer kinetic analysis method used.
SUBMITTER: Biglands JD
PROVIDER: S-EPMC4455679 | biostudies-literature | 2015 May
REPOSITORIES: biostudies-literature
Biglands John D JD Magee Derek R DR Sourbron Steven P SP Plein Sven S Greenwood John P JP Radjenovic Aleksandra A
Radiology 20141218 2
<h4>Purpose</h4>To compare the diagnostic performance of four tracer kinetic analysis methods to quantify myocardial perfusion from magnetic resonance (MR) imaging cardiac perfusion data sets in terms of their ability to lead to the diagnosis of myocardial ischemia.<h4>Materials and methods</h4>The study was approved by the regional ethics committee, and all patients gave written consent. A representative sample of 50 patients with suspected ischemic heart disease was retrospectively selected fr ...[more]