Four-year clinical follow-up of the first-in-man randomized comparison of a novel sirolimus eluting stent with abluminal biodegradable polymer and ultra-thin strut cobalt-chromium alloy: the INSPIRON-I trial.


ABSTRACT: BACKGROUND:The Inspiron™ sirolimus-eluting stent (SES) is a low-dose, ultra-thin-strut cobalt-chromium stent abluminally coated with biodegradable polymers (BP). Previous results from the INSPIRON-I trial, a first-in-man study, have proven the efficacy of the novel stent in reducing neointimal proliferation. The present report aims at evaluating the long-term clinical outcomes of patients enrolled into the INSPIRON-I trial (Clinical Trials Gov. identifier: NCT01093391). METHODS:A total of 57 patients (60 lesions) were randomly allocated in a 2:1 ratio to treatment with the Inspiron™ SES vs. its equivalent Cronus™ bare metal stent (BMS, both by Scitech Medical™, Aparecida de Goiânia, Goiás, Brazil), in four tertiary centers. The primary endpoint of the present analysis was the occurrence of major adverse cardiac events (MACE) [death, myocardial infarction (MI), target vessel revascularization (TVR) and/or target lesion revascularization (TLR)] at 4 years. RESULTS:Baseline clinical and angiographic characteristics of both groups were similar. After 4 years, the primary endpoint occurred in 7.9% vs. 23.5% of patients in Inspiron and control groups respectively (P=0.11). The rate of death/MI was similar between the groups, but there was a significant decrease in the risk of repeat revascularization in the Inspiron group compared to the control arm TLR (0.0% vs. 23.5% respectively, P=0.02). There were no stent thromboses in the study population. CONCLUSIONS:The novel Inspiron™ SES showed a sustained safe and effective clinical profile after 4-year of follow-up, with very low adverse events and null stent thrombosis (ST) occurrence.

SUBMITTER: Oliveira MD 

PROVIDER: S-EPMC4536482 | biostudies-literature | 2015 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Four-year clinical follow-up of the first-in-man randomized comparison of a novel sirolimus eluting stent with abluminal biodegradable polymer and ultra-thin strut cobalt-chromium alloy: the INSPIRON-I trial.

Oliveira Marcos Danillo Peixoto MD   Ribeiro Expedito E EE   Campos Carlos M CM   Ribeiro Henrique B HB   Faillace Bruno L R BL   Lopes Augusto C AC   Esper Rodrigo B RB   Meirelles George X GX   Perin Marco A MA   Abizaid Alexandre A   Lemos Pedro A PA  

Cardiovascular diagnosis and therapy 20150801 4


<h4>Background</h4>The Inspiron™ sirolimus-eluting stent (SES) is a low-dose, ultra-thin-strut cobalt-chromium stent abluminally coated with biodegradable polymers (BP). Previous results from the INSPIRON-I trial, a first-in-man study, have proven the efficacy of the novel stent in reducing neointimal proliferation. The present report aims at evaluating the long-term clinical outcomes of patients enrolled into the INSPIRON-I trial (Clinical Trials Gov. identifier: NCT01093391).<h4>Methods</h4>A  ...[more]

Similar Datasets

| S-EPMC3861257 | biostudies-literature
| S-EPMC8248251 | biostudies-literature
| S-EPMC7067606 | biostudies-literature
| S-EPMC4943287 | biostudies-literature
| S-EPMC6898363 | biostudies-literature
| S-EPMC6797413 | biostudies-literature
| S-EPMC5721792 | biostudies-literature
| S-EPMC7531093 | biostudies-literature
| S-EPMC10405989 | biostudies-literature