ABSTRACT: PURPOSE:MRI-guided interventions demand high frame rate imaging, making fast imaging techniques such as spiral imaging and echo planar imaging (EPI) appealing. In this study, we implemented a real-time distortion correction framework to enable the use of these fast acquisitions for interventional MRI. METHODS:Distortions caused by gradient waveform inaccuracies were corrected using the gradient impulse response function (GIRF), which was measured by standard equipment and saved as a calibration file on the host computer. This file was used at runtime to calculate the predicted k-space trajectories for image reconstruction. Additionally, the off-resonance reconstruction frequency was modified in real time to interactively deblur spiral images. RESULTS:Real-time distortion correction for arbitrary image orientations was achieved in phantoms and healthy human volunteers. The GIRF-predicted k-space trajectories matched measured k-space trajectories closely for spiral imaging. Spiral and EPI image distortion was visibly improved using the GIRF-predicted trajectories. The GIRF calibration file showed no systematic drift in 4 months and was demonstrated to correct distortions after 30 min of continuous scanning despite gradient heating. Interactive off-resonance reconstruction was used to sharpen anatomical boundaries during continuous imaging. CONCLUSIONS:This real-time distortion correction framework will enable the use of these high frame rate imaging methods for MRI-guided interventions. Magn Reson Med 75:2278-2285, 2016. © 2015 Wiley Periodicals, Inc.

SUBMITTER: Campbell-Washburn AE 

PROVIDER: S-EPMC4691439 | biostudies-literature | 2016 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Real-time distortion correction of spiral and echo planar images using the gradient system impulse response function.

Campbell-Washburn Adrienne E AE   Xue Hui H   Lederman Robert J RJ   Faranesh Anthony Z AZ   Hansen Michael S MS  

Magnetic resonance in medicine 20150626 6


<h4>Purpose</h4>MRI-guided interventions demand high frame rate imaging, making fast imaging techniques such as spiral imaging and echo planar imaging (EPI) appealing. In this study, we implemented a real-time distortion correction framework to enable the use of these fast acquisitions for interventional MRI.<h4>Methods</h4>Distortions caused by gradient waveform inaccuracies were corrected using the gradient impulse response function (GIRF), which was measured by standard equipment and saved as  ...[more]

Similar Datasets

| S-EPMC5836862 | biostudies-literature
| S-EPMC9804234 | biostudies-literature
| S-EPMC10618273 | biostudies-literature
| S-EPMC10849706 | biostudies-literature
| S-EPMC8221177 | biostudies-literature
| S-EPMC6258270 | biostudies-literature
| S-EPMC10106483 | biostudies-literature
| S-EPMC6435385 | biostudies-literature
| S-EPMC6097923 | biostudies-literature
| S-EPMC5429659 | biostudies-literature