Validation of an automated seizure detection algorithm for term neonates.


ABSTRACT: The objective of this study was to validate the performance of a seizure detection algorithm (SDA) developed by our group, on previously unseen, prolonged, unedited EEG recordings from 70 babies from 2 centres.EEGs of 70 babies (35 seizure, 35 non-seizure) were annotated for seizures by experts as the gold standard. The SDA was tested on the EEGs at a range of sensitivity settings. Annotations from the expert and SDA were compared using event and epoch based metrics. The effect of seizure duration on SDA performance was also analysed.Between sensitivity settings of 0.5 and 0.3, the algorithm achieved seizure detection rates of 52.6-75.0%, with false detection (FD) rates of 0.04-0.36FD/h for event based analysis, which was deemed to be acceptable in a clinical environment. Time based comparison of expert and SDA annotations using Cohen's Kappa Index revealed a best performing SDA threshold of 0.4 (Kappa 0.630). The SDA showed improved detection performance with longer seizures.The SDA achieved promising performance and warrants further testing in a live clinical evaluation.The SDA has the potential to improve seizure detection and provide a robust tool for comparing treatment regimens.

SUBMITTER: Mathieson SR 

PROVIDER: S-EPMC4727504 | biostudies-literature | 2016 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Validation of an automated seizure detection algorithm for term neonates.

Mathieson Sean R SR   Stevenson Nathan J NJ   Low Evonne E   Marnane William P WP   Rennie Janet M JM   Temko Andrey A   Lightbody Gordon G   Boylan Geraldine B GB  

Clinical neurophysiology : official journal of the International Federation of Clinical Neurophysiology 20150509 1


<h4>Objective</h4>The objective of this study was to validate the performance of a seizure detection algorithm (SDA) developed by our group, on previously unseen, prolonged, unedited EEG recordings from 70 babies from 2 centres.<h4>Methods</h4>EEGs of 70 babies (35 seizure, 35 non-seizure) were annotated for seizures by experts as the gold standard. The SDA was tested on the EEGs at a range of sensitivity settings. Annotations from the expert and SDA were compared using event and epoch based met  ...[more]

Similar Datasets

| S-EPMC5523521 | biostudies-other
| S-EPMC6075622 | biostudies-literature
| S-EPMC8275922 | biostudies-literature
| S-EPMC4394138 | biostudies-literature
| S-EPMC3468608 | biostudies-literature
| S-EPMC7811137 | biostudies-literature
| S-EPMC7217054 | biostudies-literature
| S-EPMC2934863 | biostudies-literature
| S-EPMC7887141 | biostudies-literature
| S-EPMC5123744 | biostudies-literature