Electromagnetic Real Time Navigation in the Region of the Posterior Pelvic Ring: An Experimental In-Vitro Feasibility Study and Comparison of Image Guided Techniques.


ABSTRACT:

Background

Electromagnetic tracking is a relatively new technique that allows real time navigation in the absence of radiation. The aim of this study was to prove the feasibility of this technique for the treatment of posterior pelvic ring fractures and to compare the results with established image guided procedures.

Methods

Tests were performed in pelvic specimens (Sawbones®) with standardized sacral fractures (Type Denis I or II). A gel matrix simulated the operative approach and a cover was used to disable visual control. The electromagnetic setup was performed by using a custom made carbon reference plate and a prototype stainless steel K-wire with an integrated sensor coil. Four different test series were performed: Group OCT: Optical navigation using preoperative CT-scans; group O3D: Optical navigation using intraoperative 3-D-fluoroscopy; group Fluoro: Conventional 2-D-fluoroscopy; group EMT: Electromagnetic navigation combined with a preoperative Dyna-CT. Accuracy of screw placement was analyzed by standardized postoperative CT-scan for each specimen. Operation time and intraoperative radiation exposure for the surgeon was documented. All data was analyzed using SPSS (Version 20, 76 Chicago, IL, USA). Statistical significance was defined as p< 0.05.

Results

160 iliosacral screws were placed (40 per group). EMT resulted in a significantly higher incidence of optimal screw placement (EMT: 36/40) compared to the groups Fluoro (30/40; p< 0.05) and OCT (31/40; p< 0.05). Results between EMT and O3D were comparable (O3D: 37/40; n.s.). Also, the operation time was comparable between groups EMT and O3D (EMT 7.62 min vs. O3D 7.98 min; n.s.), while the surgical time was significantly shorter compared to the Fluoro group (10.69 min; p< 0.001) and the OCT group (13.3 min; p< 0.001).

Conclusion

Electromagnetic guided iliosacral screw placement is a feasible procedure. In our experimental setup, this method was associated with improved accuracy of screw placement and shorter operation time when compared with the conventional fluoroscopy guided technique and compared to the optical navigation using preoperative CT-scans. Further studies are necessary to rule out drawbacks of this technique regarding ferromagnetic objects.

SUBMITTER: Pishnamaz M 

PROVIDER: S-EPMC4749384 | biostudies-literature | 2016

REPOSITORIES: biostudies-literature

altmetric image

Publications

Electromagnetic Real Time Navigation in the Region of the Posterior Pelvic Ring: An Experimental In-Vitro Feasibility Study and Comparison of Image Guided Techniques.

Pishnamaz Miguel M   Wilkmann Christoph C   Na Hong-Sik HS   Pfeffer Jochen J   Hänisch Christoph C   Janssen Max M   Bruners Philipp P   Kobbe Philipp P   Hildebrand Frank F   Schmitz-Rode Thomas T   Pape Hans-Christoph HC  

PloS one 20160210 2


<h4>Background</h4>Electromagnetic tracking is a relatively new technique that allows real time navigation in the absence of radiation. The aim of this study was to prove the feasibility of this technique for the treatment of posterior pelvic ring fractures and to compare the results with established image guided procedures.<h4>Methods</h4>Tests were performed in pelvic specimens (Sawbones®) with standardized sacral fractures (Type Denis I or II). A gel matrix simulated the operative approach an  ...[more]

Similar Datasets

| S-EPMC7330407 | biostudies-literature
| S-EPMC8339733 | biostudies-literature
| S-EPMC7996925 | biostudies-literature
| S-EPMC6339594 | biostudies-literature
| S-EPMC7797115 | biostudies-literature
| S-EPMC3563823 | biostudies-literature
| S-EPMC6344356 | biostudies-literature
| S-EPMC8107026 | biostudies-literature
| S-EPMC5351986 | biostudies-literature
| S-EPMC7309554 | biostudies-literature