Turbulent Kinetic Energy Measurement Using Phase Contrast MRI for Estimating the Post-Stenotic Pressure Drop: In Vitro Validation and Clinical Application.


ABSTRACT:

Background

Although the measurement of turbulence kinetic energy (TKE) by using magnetic resonance imaging (MRI) has been introduced as an alternative index for quantifying energy loss through the cardiac valve, experimental verification and clinical application of this parameter are still required.

Objectives

The goal of this study is to verify MRI measurements of TKE by using a phantom stenosis with particle image velocimetry (PIV) as the reference standard. In addition, the feasibility of measuring TKE with MRI is explored.

Methods

MRI measurements of TKE through a phantom stenosis was performed by using clinical 3T MRI scanner. The MRI measurements were verified experimentally by using PIV as the reference standard. In vivo application of MRI-driven TKE was explored in seven patients with aortic valve disease and one healthy volunteer. Transvalvular gradients measured by MRI and echocardiography were compared.

Results

MRI and PIV measurements of TKE are consistent for turbulent flow (0.666 < R2 < 0.738) with a mean difference of -11.13 J/m3 (SD = 4.34 J/m3). Results of MRI and PIV measurements differ by 2.76 ± 0.82 cm/s (velocity) and -11.13 ± 4.34 J/m3 (TKE) for turbulent flow (Re > 400). The turbulence pressure drop correlates strongly with total TKE (R2 = 0.986). However, in vivo measurements of TKE are not consistent with the transvalvular pressure gradient estimated by echocardiography.

Conclusions

These results suggest that TKE measurement via MRI may provide a potential benefit as an energy-loss index to characterize blood flow through the aortic valve. However, further clinical studies are necessary to reach definitive conclusions regarding this technique.

SUBMITTER: Ha H 

PROVIDER: S-EPMC4792455 | biostudies-literature | 2016

REPOSITORIES: biostudies-literature

altmetric image

Publications

Turbulent Kinetic Energy Measurement Using Phase Contrast MRI for Estimating the Post-Stenotic Pressure Drop: In Vitro Validation and Clinical Application.

Ha Hojin H   Kim Guk Bae GB   Kweon Jihoon J   Huh Hyung Kyu HK   Lee Sang Joon SJ   Koo Hyun Jung HJ   Kang Joon-Won JW   Lim Tae-Hwan TH   Kim Dae-Hee DH   Kim Young-Hak YH   Kim Namkug N   Yang Dong Hyun DH  

PloS one 20160315 3


<h4>Background</h4>Although the measurement of turbulence kinetic energy (TKE) by using magnetic resonance imaging (MRI) has been introduced as an alternative index for quantifying energy loss through the cardiac valve, experimental verification and clinical application of this parameter are still required.<h4>Objectives</h4>The goal of this study is to verify MRI measurements of TKE by using a phantom stenosis with particle image velocimetry (PIV) as the reference standard. In addition, the fea  ...[more]

Similar Datasets

| S-EPMC5397859 | biostudies-literature
| S-EPMC6899603 | biostudies-literature
| S-EPMC9513106 | biostudies-literature
| S-EPMC8620931 | biostudies-literature
| S-EPMC5556934 | biostudies-other
| S-EPMC9489053 | biostudies-literature
| S-EPMC7817331 | biostudies-literature
| S-EPMC5928300 | biostudies-literature
| S-EPMC6839367 | biostudies-literature
| S-EPMC5927556 | biostudies-literature