Unknown

Dataset Information

0

Design and testing of microbubble-based MRI contrast agents for gastric pressure measurement.


ABSTRACT:

Purpose

This work demonstrates specifically tailored microbubble-based preparations and their suitability as MRI contrast agents for ingestion and measuring temporal and spatial pressure variation in the human stomach.

Methods

Enhanced alginate spheres were prepared by incorporating gas-filled microbubbles into sodium alginate solution followed by the polymerization of the mixture in an aqueous calcium lactate solution. The microbubbles were prepared with a phospholipid shell and perfluorocarbon gas filling, using a mechanical cavitational agitation regime. The NMR signal changes to externally applied pressure and coming from the enhanced alginate spheres were acquired and compared with that of alginate spheres without microbubbles. In vivo investigations were also carried out on healthy volunteers to measure the pressure variation in the stomach.

Results

The MR signal changes in the contrast agent exhibits a linear sensitivity of approximately 40% per bar, as opposed to no measurable signal change seen in the control gas-free spheres. This novel contrast agent also demonstrates an excellent stability in simulated gastric conditions, including at body temperature. In vivo studies showed that the signal change exhibited in the meal within the antrum region is between 5% and 10%, but appears to come from both pressure changes and partial volume artifacts.

Conclusion

This study demonstrates that alginate spheres with microbubbles can be used as an MRI contrast agent to measure pressure changes. The peristaltic movement within the stomach is seen to substantially alter the overall signal intensity of the contrast agent meal. Future work must focus on improving the contrast agent's sensitivity to pressure changes.

SUBMITTER: Abdurakman E 

PROVIDER: S-EPMC6899603 | biostudies-literature | 2020 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Design and testing of microbubble-based MRI contrast agents for gastric pressure measurement.

Abdurakman Edwin E   Bencsik Martin M   Cave Gareth W V GWV   Hoad Caroline L CL   McGowan Scott S   Fairhurst David J DJ   Major Giles G   Gowland Penny A PA   Bowtell Richard R  

Magnetic resonance in medicine 20190916 3


<h4>Purpose</h4>This work demonstrates specifically tailored microbubble-based preparations and their suitability as MRI contrast agents for ingestion and measuring temporal and spatial pressure variation in the human stomach.<h4>Methods</h4>Enhanced alginate spheres were prepared by incorporating gas-filled microbubbles into sodium alginate solution followed by the polymerization of the mixture in an aqueous calcium lactate solution. The microbubbles were prepared with a phospholipid shell and  ...[more]

Similar Datasets

| S-EPMC2692035 | biostudies-literature
| S-EPMC10654417 | biostudies-literature
| S-EPMC4518216 | biostudies-literature
| S-EPMC7175143 | biostudies-literature
| S-EPMC3032015 | biostudies-literature
| S-EPMC9489053 | biostudies-literature
| S-EPMC5441847 | biostudies-literature
| S-EPMC4955692 | biostudies-literature
| S-EPMC3170541 | biostudies-literature
| S-EPMC6893250 | biostudies-literature