Unknown

Dataset Information

0

Development and evaluation of a multichannel endorectal RF coil for prostate MRI at 7T in combination with an external surface array.


ABSTRACT: To develop and evaluate a sterilizable multichannel endorectal coil (ERC) for use in combination with an external surface array (ESA) for high-resolution anatomical and functional studies of the prostate at 7T.A two-loop ERC (ERC-2L) and a microstrip-loop ERC (ERC-ML) were compared at 7T in terms of transmit and receive performance. The best-performing ERC was evaluated alone and in combination with the ESA through 1) simulations on both phantom and an anatomically correct numerical human model to assess B1+ transmit and specific absorption rate (SAR) efficiencies, and 2) phantom experiments to calculate B1+ transmit efficiency and signal-to-noise ratio (SNR). Phantom studies were also performed to look at heating when using the ERC as a transmitter and for comparing the new coil against a single-channel balloon-type ERC (ERC-b). High-resolution magnetic resonance imaging (MRI) acquisitions were performed on a single healthy subject using the two-channel ERC combined with the ESA.Compared to the ERC-ML, the ERC-2L demonstrated 20% higher SAR efficiency and higher SNR 3?cm from the coil. The presence of a tuned and detuned ERC-2L did not alter the peak local SAR of the ESA alone; however, the detuned ERC-2L had 45% less peak local SAR around the rectum compared to the tuned ERC-2L. The receive-only version of the ERC-2L improved the SNR 4.7-fold and 1.3-fold compared to the ESA and ERC-b, respectively. In combination with the ESA, the ERC-2L supported in-plane voxel-size of 0.36 × 0.36?mm(2) in T2 -weighted anatomic imaging.The reusable ERC-2L combined with an ESA offers a high SNR imaging platform for translational studies of the prostate at 7T. J. Magn. Reson. Imaging 2016;43:1279-1287.

SUBMITTER: Erturk MA 

PROVIDER: S-EPMC4865450 | biostudies-literature | 2016 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Development and evaluation of a multichannel endorectal RF coil for prostate MRI at 7T in combination with an external surface array.

Ertürk M Arcan MA   Tian Jinfeng J   Van de Moortele Pierre-François PF   Adriany Gregor G   Metzger Gregory J GJ  

Journal of magnetic resonance imaging : JMRI 20151119 6


<h4>Purpose</h4>To develop and evaluate a sterilizable multichannel endorectal coil (ERC) for use in combination with an external surface array (ESA) for high-resolution anatomical and functional studies of the prostate at 7T.<h4>Materials and methods</h4>A two-loop ERC (ERC-2L) and a microstrip-loop ERC (ERC-ML) were compared at 7T in terms of transmit and receive performance. The best-performing ERC was evaluated alone and in combination with the ESA through 1) simulations on both phantom and  ...[more]

Similar Datasets

| S-EPMC7033245 | biostudies-literature
| S-EPMC3139506 | biostudies-other
| S-EPMC3338589 | biostudies-other
| S-EPMC8575292 | biostudies-literature
| S-EPMC7309650 | biostudies-literature
| S-EPMC6899981 | biostudies-literature