Unknown

Dataset Information

0

A Novel Mono-surface Antisymmetric 8Tx/16Rx Coil Array for Parallel Transmit Cardiac MRI in Pigs at 7T.


ABSTRACT: A novel mono-surface antisymmetric 16-element transmit/receive (Tx/Rx) coil array was designed, simulated, constructed, and tested for cardiac magnetic resonance imaging (cMRI) in pigs at 7?T. The cardiac array comprised of a mono-surface 16-loops with two central elements arranged anti-symmetrically and flanked by seven elements on either side. The array was configured for parallel transmit (pTx) mode to have an eight channel transmit and 16-channel receive (8Tx/16Rx) coil array. Electromagnetic (EM) simulations, bench-top measurements, phantom, and MRI experiments with two pig cadavers (68 and 46?kg) were performed. Finally, the coil was used in pilot in-vivo measurements with a 60?kg pig. Flip angle (FA), geometry factor (g-factor), signal-to-noise ratio (SNR) maps, and high-resolution cardiac images were acquired with an in-plane resolution of 0.6?mm?×?0.6?mm (in-vivo) and 0.3?mm?×?0.3?mm (ex-vivo). The mean g-factor over the heart was 1.26 (R?=?6). Static phase [Formula: see text] shimming in a pig body phantom with the optimal phase vectors makes possible to improve the [Formula: see text] homogeneity by factor?>?2 and transmit efficiency by factor?>?3 compared to zero phases (before RF shimming). Parallel imaging performed in the in-vivo measurements demonstrated well preserved diagnostic quality of the resulting images at acceleration factors up to R?=?6. The described hardware design can be adapted for arrays optimized for animals and humans with a larger number of elements (32-64) while maintaining good decoupling for various MRI applications at UHF (e.g., cardiac, head, and spine).

SUBMITTER: Elabyad IA 

PROVIDER: S-EPMC7033245 | biostudies-literature | 2020 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

A Novel Mono-surface Antisymmetric 8Tx/16Rx Coil Array for Parallel Transmit Cardiac MRI in Pigs at 7T.

Elabyad Ibrahim A IA   Terekhov Maxim M   Lohr David D   Stefanescu Maria R MR   Baltes Steffen S   Schreiber Laura M LM  

Scientific reports 20200220 1


A novel mono-surface antisymmetric 16-element transmit/receive (Tx/Rx) coil array was designed, simulated, constructed, and tested for cardiac magnetic resonance imaging (cMRI) in pigs at 7 T. The cardiac array comprised of a mono-surface 16-loops with two central elements arranged anti-symmetrically and flanked by seven elements on either side. The array was configured for parallel transmit (pTx) mode to have an eight channel transmit and 16-channel receive (8Tx/16Rx) coil array. Electromagneti  ...[more]

Similar Datasets

| S-EPMC5947608 | biostudies-literature
| S-EPMC8423236 | biostudies-literature
| S-EPMC6618274 | biostudies-literature
| S-EPMC6107369 | biostudies-literature
| S-EPMC7047538 | biostudies-literature
| S-EPMC4761435 | biostudies-literature
| S-EPMC9315051 | biostudies-literature
| S-EPMC9314883 | biostudies-literature