Diagnostic Performance of a Smartphone-Based Photoplethysmographic Application for Atrial Fibrillation Screening in a Primary Care Setting.


ABSTRACT:

Background

Diagnosing atrial fibrillation (AF) before ischemic stroke occurs is a priority for stroke prevention in AF. Smartphone camera-based photoplethysmographic (PPG) pulse waveform measurement discriminates between different heart rhythms, but its ability to diagnose AF in real-world situations has not been adequately investigated. We sought to assess the diagnostic performance of a standalone smartphone PPG application, Cardiio Rhythm, for AF screening in primary care setting.

Methods and results

Patients with hypertension, with diabetes mellitus, and/or aged ?65 years were recruited. A single-lead ECG was recorded by using the AliveCor heart monitor with tracings reviewed subsequently by 2 cardiologists to provide the reference standard. PPG measurements were performed by using the Cardiio Rhythm smartphone application. AF was diagnosed in 28 (2.76%) of 1013 participants. The diagnostic sensitivity of the Cardiio Rhythm for AF detection was 92.9% (95% CI] 77-99%) and was higher than that of the AliveCor automated algorithm (71.4% [95% CI 51-87%]). The specificities of Cardiio Rhythm and the AliveCor automated algorithm were comparable (97.7% [95% CI: 97-99%] versus 99.4% [95% CI 99-100%]). The positive predictive value of the Cardiio Rhythm was lower than that of the AliveCor automated algorithm (53.1% [95% CI 38-67%] versus 76.9% [95% CI 56-91%]); both had a very high negative predictive value (99.8% [95% CI 99-100%] versus 99.2% [95% CI 98-100%]).

Conclusions

The Cardiio Rhythm smartphone PPG application provides an accurate and reliable means to detect AF in patients at risk of developing AF and has the potential to enable population-based screening for AF.

SUBMITTER: Chan PH 

PROVIDER: S-EPMC5015379 | biostudies-literature | 2016 Jul

REPOSITORIES: biostudies-literature

altmetric image

Publications

Diagnostic Performance of a Smartphone-Based Photoplethysmographic Application for Atrial Fibrillation Screening in a Primary Care Setting.

Chan Pak-Hei PH   Wong Chun-Ka CK   Poh Yukkee C YC   Pun Louise L   Leung Wangie Wan-Chiu WW   Wong Yu-Fai YF   Wong Michelle Man-Ying MM   Poh Ming-Zher MZ   Chu Daniel Wai-Sing DW   Siu Chung-Wah CW  

Journal of the American Heart Association 20160721 7


<h4>Background</h4>Diagnosing atrial fibrillation (AF) before ischemic stroke occurs is a priority for stroke prevention in AF. Smartphone camera-based photoplethysmographic (PPG) pulse waveform measurement discriminates between different heart rhythms, but its ability to diagnose AF in real-world situations has not been adequately investigated. We sought to assess the diagnostic performance of a standalone smartphone PPG application, Cardiio Rhythm, for AF screening in primary care setting.<h4>  ...[more]

Similar Datasets

| S-EPMC6015414 | biostudies-literature
| S-EPMC5577883 | biostudies-other
| S-EPMC6182120 | biostudies-literature
| S-EPMC9883923 | biostudies-literature
| S-EPMC5993113 | biostudies-literature
| S-EPMC7201749 | biostudies-literature
| S-EPMC10689910 | biostudies-literature
| S-EPMC8421380 | biostudies-literature
| S-EPMC8725214 | biostudies-literature
| S-EPMC7398859 | biostudies-literature