Unknown

Dataset Information

0

Coronary Computed Tomographic Angiography at 80 kVp and Knowledge-Based Iterative Model Reconstruction Is Non-Inferior to that at 100 kVp with Iterative Reconstruction.


ABSTRACT: The aims of this study were to compare the image noise and quality of coronary computed tomographic angiography (CCTA) at 80 kVp with knowledge-based iterative model reconstruction (IMR) to those of CCTA at 100 kVp with hybrid iterative reconstruction (IR), and to evaluate the feasibility of a low-dose radiation protocol with IMR. Thirty subjects who underwent prospective electrocardiogram-gating CCTA at 80 kVp, 150 mAs, and IMR (Group A), and 30 subjects with 100 kVp, 150 mAs, and hybrid IR (Group B) were retrospectively enrolled after sample-size calculation. A BMI of less than 25 kg/m2 was required for inclusion. The attenuation value and image noise of CCTA were measured and the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated at the proximal right coronary artery and left main coronary artery. The image noise was analyzed using a non-inferiority test. The CCTA images were qualitatively evaluated using a four-point scale. The radiation dose was significantly lower in Group A than Group B (0.69 ± 0.08 mSv vs. 1.39 ± 0.15 mSv, p < 0.001). The attenuation values were higher in Group A than Group B (p < 0.001). The SNR and CNR in Group A were higher than those of Group B. The image noise of Group A was non-inferior to that of Group B. Qualitative image quality of Group A was better than that of Group B (3.6 vs. 3.4, p = 0.017). CCTA at 80 kVp with IMR could reduce the radiation dose by about 50%, with non-inferior image noise and image quality than those of CCTA at 100 kVp with hybrid IR.

SUBMITTER: Lee J 

PROVIDER: S-EPMC5033462 | biostudies-literature | 2016

REPOSITORIES: biostudies-literature

altmetric image

Publications

Coronary Computed Tomographic Angiography at 80 kVp and Knowledge-Based Iterative Model Reconstruction Is Non-Inferior to that at 100 kVp with Iterative Reconstruction.

Lee Joohee J   Park Chul Hwan CH   Oh Chi Suk CS   Han Kyunghwa K   Kim Tae Hoon TH  

PloS one 20160922 9


The aims of this study were to compare the image noise and quality of coronary computed tomographic angiography (CCTA) at 80 kVp with knowledge-based iterative model reconstruction (IMR) to those of CCTA at 100 kVp with hybrid iterative reconstruction (IR), and to evaluate the feasibility of a low-dose radiation protocol with IMR. Thirty subjects who underwent prospective electrocardiogram-gating CCTA at 80 kVp, 150 mAs, and IMR (Group A), and 30 subjects with 100 kVp, 150 mAs, and hybrid IR (Gr  ...[more]

Similar Datasets

| S-EPMC6306215 | biostudies-literature
| S-EPMC5035093 | biostudies-literature
| S-EPMC7470424 | biostudies-literature
| S-EPMC9900034 | biostudies-literature
| S-EPMC9403587 | biostudies-literature
| S-EPMC10071125 | biostudies-literature
| S-EPMC5384491 | biostudies-literature
| S-EPMC5040685 | biostudies-literature
| S-EPMC6206864 | biostudies-literature
| S-EPMC5017721 | biostudies-literature