Unknown

Dataset Information

0

Dual-source abdominopelvic computed tomography: Comparison of image quality and radiation dose of 80 kVp and 80/150 kVp with tin filter.


ABSTRACT: OBJECTIVE:To compare the radiation dose and the objective and subjective image quality of 80 kVp and 80/150 kVp with tin filter (80/Sn150 kVp) computed tomography (CT) in oncology patients. METHODS:One-hundred-and-forty-five consecutive oncology patients who underwent third-generation dual-source dual-energy CT of the abdomen for evaluation of malignant visceral, peritoneal, extraperitoneal, and bone tumor were retrospectively recruited. Two radiologists independently reviewed each observation in 80 kVp CT and 80/Sn150 kVp CT. Modified line-density profile of the tumor and contrast-to-noise ratio (CNR) were measured. Diagnostic confidence, lesion conspicuity, and subjective image quality were calculated and compared between image sets. The effective dose and size-specific dose estimate (SSDE) were calculated in the image sets. RESULTS:Modified line-density profile analysis revealed higher attenuation differences between the tumor and normal tissue in 80 kVp CT than in 80/Sn150 kVp CT (127 vs. 107, P = 0.05). The 80 kVp CT showed increased CNR in the liver (8.0 vs. 7.6) and the aorta (18.9 vs. 16.3) than the 80/Sn150 kVp CT. The 80 kVp CT yielded higher enhancement of organs (4.9 ± 0.2 vs. 4.7 ± 0.4, P<0.001) and lesion conspicuity (4.9 ± 0.3 vs. 4.8 ± 0.5, P = 0.035) than the 80/Sn150 kVp CT; overall image quality and confidence index were comparable. The effective dose was reduced by 45.2% with 80 kVp CT (2.3 mSv ± 0.9) compared to 80/Sn150 kVp CT (4.1 mSv ± 1.5). The SSDE was 7.4 ± 3.8 mGy on 80/Sn150 kVp CT and 4.1 ± 2.2 mGy on 80 kVp CT. CONCLUSIONS:The 80 kVp CT reduced the radiation dose by 45.2% in oncology patients while showing comparable or superior image quality to that of 80/Sn150 kVp CT for abdominal tumor evaluation.

SUBMITTER: Choi SJ 

PROVIDER: S-EPMC7470424 | biostudies-literature | 2020

REPOSITORIES: biostudies-literature

altmetric image

Publications

Dual-source abdominopelvic computed tomography: Comparison of image quality and radiation dose of 80 kVp and 80/150 kVp with tin filter.

Choi Seung Joon SJ   Ahn Su Joa SJ   Park So Hyun SH   Park Seong Ho SH   Pak Seong Yong SY   Choi Jae Won JW   Shim Young Sup YS   Jeong Yu Mi YM   Kim Bohyun B  

PloS one 20200903 9


<h4>Objective</h4>To compare the radiation dose and the objective and subjective image quality of 80 kVp and 80/150 kVp with tin filter (80/Sn150 kVp) computed tomography (CT) in oncology patients.<h4>Methods</h4>One-hundred-and-forty-five consecutive oncology patients who underwent third-generation dual-source dual-energy CT of the abdomen for evaluation of malignant visceral, peritoneal, extraperitoneal, and bone tumor were retrospectively recruited. Two radiologists independently reviewed eac  ...[more]

Similar Datasets

| S-EPMC9403587 | biostudies-literature
| S-EPMC5033462 | biostudies-literature
| S-EPMC5040685 | biostudies-literature
| S-EPMC5035093 | biostudies-literature
| S-EPMC2077913 | biostudies-other
| S-EPMC7934306 | biostudies-literature
| S-EPMC6342763 | biostudies-literature
| S-EPMC5564335 | biostudies-literature
| S-EPMC8421356 | biostudies-literature
| S-EPMC5591158 | biostudies-literature