Unknown

Dataset Information

0

Dose-rate-dependent damage of cerium dioxide in the scanning transmission electron microscope.


ABSTRACT: Beam damage caused by energetic electrons in the transmission electron microscope is a fundamental constraint limiting the collection of artifact-free information. Through understanding the influence of the electron beam, experimental routines may be adjusted to improve the data collection process. Investigations of CeO2 indicate that there is not a critical dose required for the accumulation of electron beam damage. Instead, measurements using annular dark field scanning transmission electron microscopy and electron energy loss spectroscopy demonstrate that the onset of measurable damage occurs when a critical dose rate is exceeded. The mechanism behind this phenomenon is that oxygen vacancies created by exposure to a 300keV electron beam are actively annihilated as the sample re-oxidizes in the microscope environment. As a result, only when the rate of vacancy creation exceeds the recovery rate will beam damage begin to accumulate. This observation suggests that dose-intensive experiments can be accomplished without disrupting the native structure of the sample when executed using dose rates below the appropriate threshold. Furthermore, the presence of an encapsulating carbonaceous layer inhibits processes that cause beam damage, markedly increasing the dose rate threshold for the accumulation of damage.

SUBMITTER: Johnston-Peck AC 

PROVIDER: S-EPMC5091080 | biostudies-literature | 2016 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications

Dose-rate-dependent damage of cerium dioxide in the scanning transmission electron microscope.

Johnston-Peck Aaron C AC   DuChene Joseph S JS   Roberts Alan D AD   Wei Wei David WD   Herzing Andrew A AA  

Ultramicroscopy 20160706


Beam damage caused by energetic electrons in the transmission electron microscope is a fundamental constraint limiting the collection of artifact-free information. Through understanding the influence of the electron beam, experimental routines may be adjusted to improve the data collection process. Investigations of CeO<sub>2</sub> indicate that there is not a critical dose required for the accumulation of electron beam damage. Instead, measurements using annular dark field scanning transmission  ...[more]

Similar Datasets

| S-EPMC6392188 | biostudies-literature
| S-EPMC4758117 | biostudies-literature
| S-EPMC7033971 | biostudies-literature
| S-EPMC1413870 | biostudies-other
| S-EPMC3900039 | biostudies-literature
| S-EPMC3658130 | biostudies-literature
| S-EPMC5476802 | biostudies-literature
| S-EPMC7527223 | biostudies-literature
| S-EPMC4876401 | biostudies-literature
| S-EPMC3943919 | biostudies-literature