Calibrated bold fMRI with an optimized ASL-BOLD dual-acquisition sequence.


ABSTRACT: Calibrated fMRI techniques estimate task-induced changes in the cerebral metabolic rate of oxygen (CMRO2) based on simultaneous measurements of cerebral blood flow (CBF) and blood-oxygen-level-dependent (BOLD) signal changes evoked by stimulation. To determine the calibration factor M (corresponding to the maximum possible BOLD signal increase), BOLD signal and CBF are measured in response to a gas breathing challenge (usually CO2 or O2). Here we describe an ASL dual-acquisition sequence that combines a background-suppressed 3D-GRASE readout with 2D multi-slice EPI. The concatenation of these two imaging sequences allowed separate optimization of the acquisition for CBF and BOLD data. The dual-acquisition sequence was validated by comparison to an ASL sequence with a dual-echo EPI readout, using a visual fMRI paradigm. Results showed a 3-fold increase in temporal signal-to-noise ratio (tSNR) of the ASL time-series data while BOLD tSNR was similar to that obtained with the dual-echo sequence. The longer TR of the proposed dual-acquisition sequence, however, resulted in slightly lower T-scores (by 30%) in the BOLD activation maps. Further, the potential of the dual-acquisition sequence for M-mapping on the basis of a hypercapnia gas breathing challenge and for quantification of CMRO2 changes in response to a motor activation task was assessed. In five subjects, an average gray matter M-value of 8.71±1.03 and fractional changes of CMRO2 of 12.5±5% were found. The new sequence remedies the deficiencies of prior combined BOLD-ASL acquisition strategies by substantially enhancing perfusion tSNR, which is essential for accurate BOLD calibration.

SUBMITTER: Fernandez-Seara MA 

PROVIDER: S-EPMC5159288 | biostudies-literature | 2016 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications

Calibrated bold fMRI with an optimized ASL-BOLD dual-acquisition sequence.

Fernández-Seara María A MA   Rodgers Zachary B ZB   Englund Erin K EK   Wehrli Felix W FW  

NeuroImage 20160805


Calibrated fMRI techniques estimate task-induced changes in the cerebral metabolic rate of oxygen (CMRO<sub>2</sub>) based on simultaneous measurements of cerebral blood flow (CBF) and blood-oxygen-level-dependent (BOLD) signal changes evoked by stimulation. To determine the calibration factor M (corresponding to the maximum possible BOLD signal increase), BOLD signal and CBF are measured in response to a gas breathing challenge (usually CO<sub>2</sub> or O<sub>2</sub>). Here we describe an ASL  ...[more]

Similar Datasets

| S-EPMC9120552 | biostudies-literature
| S-EPMC4435663 | biostudies-literature
| S-EPMC4733593 | biostudies-literature
| S-EPMC6264385 | biostudies-other
| S-EPMC5448847 | biostudies-other
| S-EPMC7722021 | biostudies-literature
| S-EPMC4262662 | biostudies-literature
| S-EPMC6248456 | biostudies-literature
| S-EPMC8545051 | biostudies-literature
| S-EPMC10962577 | biostudies-literature