Performance of FACSPresto Point-of-Care Instrument for CD4-T Cell Enumeration in Human Immunodeficiency Virus (HIV)-Infected Patients Attending Care and Treatment Clinics in Belgium and Tanzania.


ABSTRACT:

Background

CD4 T-cell counts are widely used to assess treatment eligibility and to follow-up HIV-infected patients. The World Health Organization prequalification of in vitro diagnostics program conducted a performance evaluation of the FACSPresto (BD Biosciences), a new point-of-care instrument to measure absolute CD4-T cell (CD4) counts and percentages in venous and capillary blood samples from HIV-infected patients.

Methods

Patients were recruited in Belgium (200 patients) and in Tanzania (247 patients). Venous blood samples were analyzed in two nearby reference laboratories. In addition, nurses/technicians collected a capillary blood sample by finger prick directly into a FACSPresto CD4 cartridge. Assay precision was assessed on fresh blood and on external quality control samples. Trueness (bias) was assessed by comparing results from FACSPresto with the reference (single-platform FACSCalibur). Clinical misclassification was measured at 200, 350 and 500 cells/?L thresholds.

Results

Intra-assay precision was < 6%, and inter-assay < 8%. CD4 results from FACSPresto and reference method resulted in regression slopes of 0.99-1.11 using either venous or capillary blood. Correlation was better for venous than for capillary blood (minimum 0.97 vs 0.93 respectively). Capillary blood resulted in a larger bias than venous blood, with 24 and 83 cells/?L for absolute CD4 counts on capillary blood in Antwerp and Dar es Salaam respectively, vs 12 and 41 cells/?L on venous blood. Bias on CD4% was < 1% on both venous and capillary blood, and was proportionally better than for absolute CD4 counts. Clinical misclassification was in line with the average overestimation, showing a very good specificity, but sensitivity around 70-90%. The rejection rate was 11% on first reading, leading to 6% of all samples without final result after a second reading.

Conclusions

The FACSPresto performed very well on venous blood samples, and well on capillary blood samples.

SUBMITTER: Daneau G 

PROVIDER: S-EPMC5271305 | biostudies-literature | 2017

REPOSITORIES: biostudies-literature

altmetric image

Publications

Performance of FACSPresto Point-of-Care Instrument for CD4-T Cell Enumeration in Human Immunodeficiency Virus (HIV)-Infected Patients Attending Care and Treatment Clinics in Belgium and Tanzania.

Daneau Géraldine G   Aboud Said S   Prat Irena I   Urassa Willy W   Kestens Luc L  

PloS one 20170127 1


<h4>Background</h4>CD4 T-cell counts are widely used to assess treatment eligibility and to follow-up HIV-infected patients. The World Health Organization prequalification of in vitro diagnostics program conducted a performance evaluation of the FACSPresto (BD Biosciences), a new point-of-care instrument to measure absolute CD4-T cell (CD4) counts and percentages in venous and capillary blood samples from HIV-infected patients.<h4>Methods</h4>Patients were recruited in Belgium (200 patients) and  ...[more]

Similar Datasets

| S-EPMC6881989 | biostudies-literature
| S-EPMC10801968 | biostudies-literature
| S-EPMC10653711 | biostudies-literature
| S-EPMC6415788 | biostudies-literature
| S-EPMC8941073 | biostudies-literature
| S-EPMC4955015 | biostudies-literature
| S-EPMC4693083 | biostudies-literature
| 2404133 | ecrin-mdr-crc
| S-EPMC6355822 | biostudies-literature