Biomechanical investigation of two plating systems for medial column fusion in foot.


ABSTRACT:

Background

Arthrodesis of the medial column (navicular, cuneiform I and metatarsal I) is performed for reasons such as Charcot arthropathy, arthritis, posttraumatic reconstruction or severe pes planus. However, the complication rate is still high and mainly resulting from inadequate fixation. Special plates, designed for medial column arthrodesis, seem to offer potential to reduce the complication rate. The aim of this study was to investigate biomechanically plantar and dorsomedial fusion of the medial column using two new plating systems.

Methods

Eight matched pairs of human cadaveric lower legs were randomized in two groups and medial column fusion was performed using either plantar or dorsomedial variable-angle locking compression plates. The specimens were biomechanically tested under cyclic progressively increasing axial loading with physiological profile of each cycle. In addition to the machine data, mediolateral x-rays were taken every 250 cycles and motion tracking was performed to determine movements at the arthrodesis site. Statistical analysis of the parameters of interest was performed at a level of significance p = 0.05.

Results

Displacement of the talo-navicular joint after 1000, 2000 and 4000 cycles was significantly lower for plantar plating (p?0.039) while there was significantly less movement in the naviculo-cuneiform I joint for dorsal plating post these cycle numbers (p<0.001). Displacements in all three joints of the medial column, as well as angular and torsional deformations between the navicular and metatarsal I increased significantly for each plating technique between 1000, 2000 and 4000 cycles (p?0.021). The two plating systems did not differ significantly with regard to stiffness and cycles to failure (p?0.171).

Conclusion

From biomechanical point of view, although dorsomedial plating showed less movement than plantar plating in the current setup under dynamic loading, there was no significant difference between the two plating systems with regard to stiffness and cycles to failure. Both tested techniques for dorsomedial and plantar plating appear to be applicable for arthrodesis of the medial column of the foot and other considerations, such as access morbidity, associated deformities or surgeon's preference, may also guide the choice of plating pattern. Further clinical studies are necessary before definitive recommendations can be given.

SUBMITTER: Simons P 

PROVIDER: S-EPMC5319781 | biostudies-literature | 2017

REPOSITORIES: biostudies-literature

altmetric image

Publications

Biomechanical investigation of two plating systems for medial column fusion in foot.

Simons Paul P   Sommerer Theresia T   Zderic Ivan I   Wahl Dieter D   Lenz Mark M   Skulev Hristo H   Knobe Matthias M   Gueorguiev Boyko B   Richards R Geoff RG   Klos Kajetan K  

PloS one 20170221 2


<h4>Background</h4>Arthrodesis of the medial column (navicular, cuneiform I and metatarsal I) is performed for reasons such as Charcot arthropathy, arthritis, posttraumatic reconstruction or severe pes planus. However, the complication rate is still high and mainly resulting from inadequate fixation. Special plates, designed for medial column arthrodesis, seem to offer potential to reduce the complication rate. The aim of this study was to investigate biomechanically plantar and dorsomedial fusi  ...[more]

Similar Datasets

| S-EPMC4038905 | biostudies-literature
| S-EPMC9775186 | biostudies-literature
2005-12-20 | GSE2595 | GEO
| S-EPMC8220611 | biostudies-literature
| S-EPMC6739430 | biostudies-literature
2009-09-23 | GSE18179 | GEO
2009-09-25 | E-GEOD-18179 | biostudies-arrayexpress
| S-EPMC5695793 | biostudies-literature
| S-EPMC8339138 | biostudies-literature
| S-EPMC3431260 | biostudies-literature