A Clinical Trial of the Accuracy and Treatment Experience of the Flash Glucose Monitor FreeStyle Libre in Adults with Type 1 Diabetes.


ABSTRACT: In Sweden, FreeStyle Libre a flash glucose monitoring system came onto the market in 2014 as a complement to self-monitoring of blood glucose. The aim of this study was to evaluate the accuracy and treatment experience of the FreeStyle Libre system.Fifty-eight adults with type 1 diabetes used FreeStyle Libre for 10-14 days and measured capillary blood glucose levels with the HemoCue blood glucose measurement system at least six times a day simultaneously.For the entire study period, the mean absolute relative difference (MARD) was 13.2% (95% confidence interval [CI] 12.0%-14.4%). MARD was 13.6% (95% CI 12.1%-15.4%) during week 1 and 12.7% (95% CI 11.5%-13.9%) during week 2. The mean absolute difference (MAD) for the whole study period was 19.8?mg/dL (1.1?mmol/L) (95% CI 17.8-21.8?mg/dL), including 20.5?mg/dL (1.14?mmol/L) during week 1 and 19.0?mg/dL (1.05?mmol/L) during week 2. The overall correlation coefficient was 0.96. For glucose values <72, 72-180, and >180?mg/dL (<4, 4-10, and >10?mmol/L), the MARD was 20.3% (95% CI 17.7%-23.1%), 14.7% (95% CI 13.4%-16%), and 9.6% (95% CI 8.5%-10.8%), respectively, and respective MAD values were 12.3, 17.8, and 23.6?mg/dL (0.69, 0.99, and 1.31?mmol/L). Using the 10-item visual analog scale, patients rated their experience with FreeStyle Libre as generally positive, with mean values ranging from 8.22 to 9.8.FreeStyle Libre had a similar overall MARD as continuous blood glucose monitoring systems in earlier studies when studied in similar at-home conditions. The overall patient satisfaction was high.

SUBMITTER: Olafsdottir AF 

PROVIDER: S-EPMC5359691 | biostudies-literature | 2017 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

A Clinical Trial of the Accuracy and Treatment Experience of the Flash Glucose Monitor FreeStyle Libre in Adults with Type 1 Diabetes.

Ólafsdóttir Arndís F AF   Attvall Stig S   Sandgren Ulrika U   Dahlqvist Sofia S   Pivodic Aldina A   Skrtic Stanko S   Theodorsson Elvar E   Lind Marcus M  

Diabetes technology & therapeutics 20170306 3


<h4>Background</h4>In Sweden, FreeStyle Libre a flash glucose monitoring system came onto the market in 2014 as a complement to self-monitoring of blood glucose. The aim of this study was to evaluate the accuracy and treatment experience of the FreeStyle Libre system.<h4>Methods</h4>Fifty-eight adults with type 1 diabetes used FreeStyle Libre for 10-14 days and measured capillary blood glucose levels with the HemoCue blood glucose measurement system at least six times a day simultaneously.<h4>Re  ...[more]

Similar Datasets

| S-EPMC9315169 | biostudies-literature
| S-EPMC7578738 | biostudies-literature
| S-EPMC8280849 | biostudies-literature
| S-EPMC6821586 | biostudies-literature
| S-EPMC6875221 | biostudies-literature
| S-EPMC7574406 | biostudies-literature
| S-EPMC7809713 | biostudies-literature
| S-EPMC10241748 | biostudies-literature
| S-EPMC6354746 | biostudies-literature
| S-EPMC5887912 | biostudies-literature