A directional 103Pd brachytherapy device: Dosimetric characterization and practical aspects for clinical use.


ABSTRACT: A brachytherapy (BT) device has been developed with shielding to provide directional BT for preferentially irradiating malignancies while sparing healthy tissues. The CivaSheet is a flexible low-dose-rate BT device containing CivaDots with 103Pd shielded by a thin Au disk. This is the first report of a clinical dosimetric characterization of the CivaSheet device.Radiation dose distributions near a CivaDot were estimated using the MCNP6 radiation transport code. CivaSheet arrays were also modeled to evaluate the dose superposition principle for treatment planning. The resultant data were commissioned in a treatment planning system (TPS) (VariSeed 9.0), and the accuracy of the dose superposition principle was evaluated for summing individual elements comprising a planar CivaSheet.The dose-rate constant (0.579 cGy/h/U) was lower than for 103Pd seeds due to Au L-shell x-rays increasing the air-kerma strength. Radial dose function values at 0.1, 0.5, 2, 5, and 10 cm were 1.884, 1.344, 0.558, 0.088, and 0.0046, respectively. The two-dimensional anisotropy function exhibited dramatic reduction between the forward (0°) and rearward (180°) directions by a factor of 276 at r = 0.1 cm, 24 at r = 1 cm, and 5.3 at r = 10 cm. This effect diminished due to increasingly scattered radiation. The largest gradient in the two-dimensional anisotropy function was in contact with the device at 92° due to the Au disk shielding. TPS commissioning and dose superposition accuracies were typically within 2%.Simulations of the CivaDot yielded comprehensive dosimetry parameters that were entered into a TPS and deemed acceptable for clinical use. Dosimetry measurements of the CivaSheet are also of interest to the BT community.

SUBMITTER: Rivard MJ 

PROVIDER: S-EPMC5362288 | biostudies-literature | 2017 Mar - Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

A directional <sup>103</sup>Pd brachytherapy device: Dosimetric characterization and practical aspects for clinical use.

Rivard Mark J MJ  

Brachytherapy 20161228 2


<h4>Purpose</h4>A brachytherapy (BT) device has been developed with shielding to provide directional BT for preferentially irradiating malignancies while sparing healthy tissues. The CivaSheet is a flexible low-dose-rate BT device containing CivaDots with <sup>103</sup>Pd shielded by a thin Au disk. This is the first report of a clinical dosimetric characterization of the CivaSheet device.<h4>Methods and materials</h4>Radiation dose distributions near a CivaDot were estimated using the MCNP6 rad  ...[more]

Similar Datasets

| S-EPMC6548702 | biostudies-literature
| S-EPMC9144752 | biostudies-literature
| S-EPMC5833894 | biostudies-literature
| S-EPMC5722760 | biostudies-literature
| S-EPMC10443333 | biostudies-literature
| S-EPMC8335895 | biostudies-literature
| S-EPMC3987645 | biostudies-other
| S-EPMC5722622 | biostudies-literature
2013-04-24 | E-GEOD-42265 | biostudies-arrayexpress
| S-EPMC6234692 | biostudies-literature