Determination of the Porosity of PLGA Microparticles by Tracking Their Sedimentation Velocity Using a Flow Imaging Microscope (FlowCAM).


ABSTRACT:

Purpose

To investigate whether particle sedimentation velocity tracking using a flow imaging microscope (FlowCAM) can be used to determine microparticle porosity.

Methods

Two different methods were explored. In the first method the sedimentation rate of microparticles was tracked in suspending media with different densities. The porosity was calculated from the average apparent density of the particles derived by inter- or extrapolation to the density of a suspending medium in which the sedimentation velocity was zero. In the second method, the microparticle size and sedimentation velocity in one suspending fluid were used to calculate the density and porosity of individual particles by using the Stokes' law of sedimentation.

Results

Polystyrene beads of different sizes were used for the development, optimization and validation of the methods. For both methods we found porosity values that were in excellent agreement with the expected values. Both methods were applied to determine the porosity of three PLGA microparticle batches with different porosities (between about 4 and 52%). With both methods we obtained microparticle porosity values similar to those obtained by mercury intrusion porosimetry.

Conclusions

We developed two methods to determine average microparticle density and porosity by sedimentation velocity tracking, using only a few milligrams of powder.

SUBMITTER: Sediq AS 

PROVIDER: S-EPMC5382175 | biostudies-literature | 2017 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

Determination of the Porosity of PLGA Microparticles by Tracking Their Sedimentation Velocity Using a Flow Imaging Microscope (FlowCAM).

Sediq A S AS   Waasdorp S K D SKD   Nejadnik M R MR   van Beers M M C MMC   Meulenaar J J   Verrijk R R   Jiskoot W W  

Pharmaceutical research 20170217 5


<h4>Purpose</h4>To investigate whether particle sedimentation velocity tracking using a flow imaging microscope (FlowCAM) can be used to determine microparticle porosity.<h4>Methods</h4>Two different methods were explored. In the first method the sedimentation rate of microparticles was tracked in suspending media with different densities. The porosity was calculated from the average apparent density of the particles derived by inter- or extrapolation to the density of a suspending medium in whi  ...[more]

Similar Datasets

| S-EPMC3089910 | biostudies-literature
| S-EPMC3640771 | biostudies-literature
| S-EPMC3058617 | biostudies-literature
| S-EPMC6158090 | biostudies-literature
| S-EPMC5880337 | biostudies-literature
| S-EPMC4850477 | biostudies-literature
| S-EPMC5899733 | biostudies-literature
| S-EPMC10503714 | biostudies-literature
| S-EPMC8098776 | biostudies-literature
| S-EPMC5543194 | biostudies-literature