18F-FDG silicon photomultiplier PET/CT: A pilot study comparing semi-quantitative measurements with standard PET/CT.


ABSTRACT:

Purpose

To evaluate if the new Discovery Molecular Insights (DMI) PET/CT scanner provides equivalent results compared to the standard of care PET/CT scanners (GE Discovery 600 or GE Discovery 690) used in our clinic and to explore any possible differences in semi-quantitative measurements.

Methods

The local Institutional Review Board approved the protocol and written informed consent was obtained from each patient. Between September and November 2016, 50 patients underwent a single 18F-FDG injection and two scans: the clinical standard PET/CT followed immediately by the DMI PET/CT scan. We measured SUVmax and SUVmean of different background organs and up to four lesions per patient from data acquired using both scanners.

Results

DMI PET/CT identified all the 107 lesions detected by standard PET/CT scanners, as well as additional 37 areas of focal increased 18F-FDG uptake. The SUVmax values for all 107 lesions ranged 1.2 to 14.6 (mean ± SD: 2.8 ± 2.8), higher on DMI PET/CT compared with standard of care PET/CT. The mean lesion:aortic arch SUVmax ratio and mean lesion:liver SUVmax ratio were 0.2-15.2 (mean ± SD: 3.2 ± 2.6) and 0.2-8.5 (mean ± SD: 1.9 ± 1.4) respectively, higher on DMI PET/CT than standard PET/CT. These differences were statistically significant (P value < 0.0001) and not correlated to the delay in acquisition of DMI PET data (P < 0.0001).

Conclusions

Our study shows high performance of the new DMI PET/CT scanner. This may have a significant role in diagnosing and staging disease, as well as for assessing and monitoring responses to therapies.

SUBMITTER: Baratto L 

PROVIDER: S-EPMC5459477 | biostudies-literature | 2017

REPOSITORIES: biostudies-literature

altmetric image

Publications

18F-FDG silicon photomultiplier PET/CT: A pilot study comparing semi-quantitative measurements with standard PET/CT.

Baratto Lucia L   Park Sonya Young SY   Hatami Negin N   Davidzon Guido G   Srinivas Shyam S   Gambhir Sanjiv Sam SS   Iagaru Andrei A  

PloS one 20170605 6


<h4>Purpose</h4>To evaluate if the new Discovery Molecular Insights (DMI) PET/CT scanner provides equivalent results compared to the standard of care PET/CT scanners (GE Discovery 600 or GE Discovery 690) used in our clinic and to explore any possible differences in semi-quantitative measurements.<h4>Methods</h4>The local Institutional Review Board approved the protocol and written informed consent was obtained from each patient. Between September and November 2016, 50 patients underwent a singl  ...[more]

Similar Datasets

| S-EPMC8410931 | biostudies-literature
| S-EPMC5221689 | biostudies-other
| S-EPMC8803713 | biostudies-literature
| S-EPMC5987871 | biostudies-other
| S-EPMC7010890 | biostudies-literature
| S-EPMC7952491 | biostudies-literature
| S-EPMC8410733 | biostudies-literature
| S-EPMC7937050 | biostudies-literature
| S-EPMC6915187 | biostudies-literature
| S-EPMC5445063 | biostudies-literature