Performance analysis of a new hypersonic vitrector system.


ABSTRACT:

Purpose

To evaluate porcine vitreous flow and water flow rates in a new prototype hypersonic vitrectomy system compared to currently available pneumatic guillotine vitrectors (GVs) systems.

Methods

Two vitrectors were tested, a prototype, ultrasound-powered, hypersonic vitrector (HV) and a GV. Porcine vitreous was obtained within 12 to 24 h of sacrifice and kept at 4°C. A vial of vitreous or water was placed on a precision balance and its weight measured before and after the use of each vitrector. Test parameters included changes in aspiration levels, vitrector gauge, cut rates for GVs, % ultrasound (US) power for HVs, and port size for HVs. Data was analysed using linear regression and t-tests.

Results

There was no difference in the total average mean water flow between the 25-gauge GV and the 25-gauge HV (t-test: P = 0.363); however, 25-gauge GV was superior (t-test: P < 0.001) in vitreous flow. The 23-gauge GV was only more efficient in water and vitreous removal than 23-gauge HV needle-1 (Port 0.0055) (t-test: P < 0.001). For HV, wall thickness and gauge had no effect on flow rates. Water and vitreous flows showed a direct correlation with increasing aspiration levels and % US power (p<0.05).

Conclusions

The HV produced consistent water and vitreous flow rates across the range of US power and aspiration levels tested. Hypersonic vitrectomy may be a promising new alternative to the currently available guillotine-based technologies.

SUBMITTER: Stanga PE 

PROVIDER: S-EPMC5460816 | biostudies-literature | 2017

REPOSITORIES: biostudies-literature

altmetric image

Publications

Performance analysis of a new hypersonic vitrector system.

Stanga Paulo Eduardo PE   Pastor-Idoate Salvador S   Zambrano Isaac I   Carlin Paul P   McLeod David D  

PloS one 20170606 6


<h4>Purpose</h4>To evaluate porcine vitreous flow and water flow rates in a new prototype hypersonic vitrectomy system compared to currently available pneumatic guillotine vitrectors (GVs) systems.<h4>Methods</h4>Two vitrectors were tested, a prototype, ultrasound-powered, hypersonic vitrector (HV) and a GV. Porcine vitreous was obtained within 12 to 24 h of sacrifice and kept at 4°C. A vial of vitreous or water was placed on a precision balance and its weight measured before and after the use o  ...[more]

Similar Datasets

| S-EPMC4595630 | biostudies-literature
| S-EPMC8157988 | biostudies-literature
| S-EPMC6544375 | biostudies-literature
| S-EPMC8515926 | biostudies-literature
| S-EPMC4851006 | biostudies-literature
| S-EPMC6755568 | biostudies-literature
| S-EPMC4129174 | biostudies-other
| S-EPMC5835467 | biostudies-literature
| S-EPMC6251131 | biostudies-other
| S-EPMC7374231 | biostudies-literature