Unknown

Dataset Information

0

Simultaneous multislice imaging for native myocardial T1 mapping: Improved spatial coverage in a single breath-hold.


ABSTRACT: To develop a saturation recovery myocardial T1 mapping method for the simultaneous multislice acquisition of three slices.Saturation pulse-prepared heart rate independent inversion recovery (SAPPHIRE) T1 mapping was implemented with simultaneous multislice imaging using FLASH readouts for faster coverage of the myocardium. Controlled aliasing in parallel imaging (CAIPI) was used to achieve minimal noise amplification in three slices. Multiband reconstruction was performed using three linear reconstruction methods: Slice- and in-plane GRAPPA, CG-SENSE, and Tikhonov-regularized CG-SENSE. Accuracy, spatial variability, and interslice leakage were compared with single-band T1 mapping in a phantom and in six healthy subjects.Multiband phantom T1 times showed good agreement with single-band T1 mapping for all three reconstruction methods (normalized root mean square error <1.0%). The increase in spatial variability compared with single-band imaging was lowest for GRAPPA (1.29-fold), with higher penalties for Tikhonov-regularized CG-SENSE (1.47-fold) and CG-SENSE (1.52-fold). In vivo multiband T1 times showed no significant difference compared with single-band (T1 time?±?intersegmental variability: single-band, 1580?±?119 ms; GRAPPA, 1572?±?145 ms; CG-SENSE, 1579?±?159 ms; Tikhonov, 1586?±?150 ms [analysis of variance; P?=?0.86]). Interslice leakage was smallest for GRAPPA (5.4%) and higher for CG-SENSE (6.2%) and Tikhonov-regularized CG-SENSE (7.9%).Multiband accelerated myocardial T1 mapping demonstrated the potential for single-breath-hold T1 quantification in 16 American Heart Association segments over three slices. A 1.2- to 1.4-fold higher in vivo spatial variability was observed, where GRAPPA-based reconstruction showed the highest homogeneity and the least interslice leakage. Magn Reson Med 78:462-471, 2017. © 2017 International Society for Magnetic Resonance in Medicine.

SUBMITTER: Weingartner S 

PROVIDER: S-EPMC5509494 | biostudies-literature | 2017 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Simultaneous multislice imaging for native myocardial T<sub>1</sub> mapping: Improved spatial coverage in a single breath-hold.

Weingärtner Sebastian S   Moeller Steen S   Schmitter Sebastian S   Auerbach Edward E   Kellman Peter P   Shenoy Chetan C   Akçakaya Mehmet M  

Magnetic resonance in medicine 20170605 2


<h4>Purpose</h4>To develop a saturation recovery myocardial T<sub>1</sub> mapping method for the simultaneous multislice acquisition of three slices.<h4>Methods</h4>Saturation pulse-prepared heart rate independent inversion recovery (SAPPHIRE) T<sub>1</sub> mapping was implemented with simultaneous multislice imaging using FLASH readouts for faster coverage of the myocardium. Controlled aliasing in parallel imaging (CAIPI) was used to achieve minimal noise amplification in three slices. Multiban  ...[more]

Similar Datasets

| S-EPMC5980781 | biostudies-literature
| S-EPMC7669807 | biostudies-literature
| S-EPMC5871611 | biostudies-literature
| S-EPMC4864326 | biostudies-other
| S-EPMC4786209 | biostudies-literature
| S-EPMC5963299 | biostudies-literature
| S-EPMC7977704 | biostudies-literature