Reliable clinical serum analysis with reusable electrochemical sensor: Toward point-of-care measurement of the antipsychotic medication clozapine.


ABSTRACT: Clozapine is one of the most promising medications for managing schizophrenia but it is under-utilized because of the challenges of maintaining serum levels in a safe therapeutic range (1-3?M). Timely measurement of serum clozapine levels has been identified as a barrier to the broader use of clozapine, which is however challenging due to the complexity of serum samples. We demonstrate a robust and reusable electrochemical sensor with graphene-chitosan composite for rapidly measuring serum levels of clozapine. Our electrochemical measurements in clinical serum from clozapine-treated and clozapine-untreated schizophrenia groups are well correlated to centralized laboratory analysis for the readily detected uric acid and for the clozapine which is present at 100-fold lower concentration. The benefits of our electrochemical measurement approach for serum clozapine monitoring are: (i) rapid measurement (?20min) without serum pretreatment; (ii) appropriate selectivity and sensitivity (limit of detection 0.7?M); (iii) reusability of an electrode over several weeks; and (iv) rapid reliability testing to detect common error-causing problems. This simple and rapid electrochemical approach for serum clozapine measurements should provide clinicians with the timely point-of-care information required to adjust dosages and personalize the management of schizophrenia.

SUBMITTER: Kang M 

PROVIDER: S-EPMC5523846 | biostudies-literature | 2017 Sep

REPOSITORIES: biostudies-literature

altmetric image

Publications

Reliable clinical serum analysis with reusable electrochemical sensor: Toward point-of-care measurement of the antipsychotic medication clozapine.

Kang Mijeong M   Kim Eunkyoung E   Winkler Thomas E TE   Banis George G   Liu Yi Y   Kitchen Christopher A CA   Kelly Deanna L DL   Ghodssi Reza R   Payne Gregory F GF  

Biosensors & bioelectronics 20170412


Clozapine is one of the most promising medications for managing schizophrenia but it is under-utilized because of the challenges of maintaining serum levels in a safe therapeutic range (1-3μM). Timely measurement of serum clozapine levels has been identified as a barrier to the broader use of clozapine, which is however challenging due to the complexity of serum samples. We demonstrate a robust and reusable electrochemical sensor with graphene-chitosan composite for rapidly measuring serum level  ...[more]

Similar Datasets

| S-EPMC1449638 | biostudies-literature
| S-EPMC3894721 | biostudies-other
| S-EPMC6344552 | biostudies-literature
| S-EPMC4902215 | biostudies-literature
| S-EPMC5509641 | biostudies-literature
| S-EPMC8097980 | biostudies-literature
| S-EPMC6464566 | biostudies-literature
| S-EPMC5469345 | biostudies-literature
| S-EPMC4105695 | biostudies-other
| S-EPMC5241045 | biostudies-literature