Human Thiel-Embalmed Cadaveric Aortic Model with Perfusion for Endovascular Intervention Training and Medical Device Evaluation.


ABSTRACT:

Purpose

The purpose of this investigation was to evaluate human Thiel-embalmed cadavers with the addition of extracorporeal driven ante-grade pulsatile flow in the aorta as a model for simulation training in interventional techniques and endovascular device testing.

Materials and methods

Three human cadavers embalmed according to the method of Thiel were selected. Extracorporeal pulsatile ante-grade flow of 2.5 L per min was delivered directly into the aorta of the cadavers via a surgically placed connection. During perfusion, aortic pressure and temperature were recorded and optimized for physiologically similar parameters. Pre- and post-procedure CT imaging was conducted to plan and follow up thoracic and abdominal endovascular aortic repair as it would be in a clinical scenario. Thoracic endovascular aortic repair (TEVAR) and endovascular abdominal repair (EVAR) procedures were conducted in simulation of a clinical case, under fluoroscopic guidance with a multidisciplinary team present.

Results

The Thiel cadaveric aortic perfusion model provided pulsatile ante-grade flow, with pressure and temperature, sufficient to conduct a realistic simulation of TEVAR and EVAR procedures. Fluoroscopic imaging provided guidance during the intervention. Pre- and post-procedure CT imaging facilitated planning and follow-up evaluation of the procedure.

Conclusion

The human Thiel-embalmed cadavers with the addition of extracorporeal flow within the aorta offer an anatomically appropriate, physiologically similar robust model to simulate aortic endovascular procedures, with potential applications in interventional radiology training and medical device testing as a pre-clinical model.

SUBMITTER: McLeod H 

PROVIDER: S-EPMC5541076 | biostudies-literature | 2017 Sep

REPOSITORIES: biostudies-literature

altmetric image

Publications

Human Thiel-Embalmed Cadaveric Aortic Model with Perfusion for Endovascular Intervention Training and Medical Device Evaluation.

McLeod Helen H   Cox Ben F BF   Robertson James J   Duncan Robyn R   Matthew Shona S   Bhat Raj R   Barclay Avril A   Anwar J J   Wilkinson Tracey T   Melzer Andreas A   Houston J Graeme JG  

Cardiovascular and interventional radiology 20170427 9


<h4>Purpose</h4>The purpose of this investigation was to evaluate human Thiel-embalmed cadavers with the addition of extracorporeal driven ante-grade pulsatile flow in the aorta as a model for simulation training in interventional techniques and endovascular device testing.<h4>Materials and methods</h4>Three human cadavers embalmed according to the method of Thiel were selected. Extracorporeal pulsatile ante-grade flow of 2.5 L per min was delivered directly into the aorta of the cadavers via a  ...[more]

Similar Datasets

| S-EPMC9167811 | biostudies-literature
| S-EPMC8173723 | biostudies-literature
2022-08-21 | GSE197756 | GEO
| S-EPMC6944262 | biostudies-literature
| PRJNA812000 | ENA
| S-EPMC8556492 | biostudies-literature
| S-EPMC10071004 | biostudies-literature
| S-EPMC8077032 | biostudies-literature
| S-EPMC5483202 | biostudies-other
| S-EPMC11372803 | biostudies-literature