The local clinical validation of a new lithium heparin tube with a barrier: BD Vacutainer® Barricor LH Plasma tube.


ABSTRACT:

Introduction

Although serum-providing blood tubes with a barrier are still widely used due to their significant advantages, the use of blood tubes with a barrier to provide plasma is becoming widespread. We compared 22 analytes in a BD Vacutainer® Barricor LH Plasma tube for local clinical validation of this new lithium heparin tube with a barrier.

Materials and methods

Samples from 44 volunteers were collected in different tubes (Becton Dickinson and Company): Z tube without additive (reference), clot-activator tube with gel (SST), lithium heparin tube without gel (LiH), and lithium heparin tube with barrier (Barricor). Analyte concentrations in different tubes were compared with the reference tube. All tubes were also evaluated according to additional testing (different centrifugation durations, blood-sampling techniques and individual differences).

Results

Aspartate aminotransferase (AST), glucose (Glc), potassium (K), lactate dehydrogenase (LD), sodium (Na), and total protein (TP) had a significant bias in Barricor (9.19%, - 3.24%, - 4.88%, 21.60%, - 0.40%, 5.03%, respectively) relative to the reference tube. There was no statistical difference between different centrifugation durations and individual differences for AST, K and LD in LiH and/or Barricor (P > 0.05). There was a significant bias for LD between LiH and Barricor in terms of blood-sampling techniques (21.2% and 12.4%, respectively).

Conclusions

Recently, the use of plasma has become prominent due to some of its advantages. In this study, plasma AST, K, LD, Glc and TP levels in Barricor were clinically different in comparison to serum. The results of additional tests showed that higher levels of LD in Barricor did not result from haemolysis, and they might be related to other factors including number of platelets, cellular fragility, or functional environment.

SUBMITTER: Arslan FD 

PROVIDER: S-EPMC5575652 | biostudies-literature | 2017 Oct

REPOSITORIES: biostudies-literature

altmetric image

Publications

The local clinical validation of a new lithium heparin tube with a barrier: BD Vacutainer® Barricor LH Plasma tube.

Arslan Fatma Demet FD   Karakoyun Inanc I   Basok Banu Isbilen BI   Aksit Merve Zeytinli MZ   Baysoy Anil A   Ozturk Yasemin Kilic YK   Guclu Yusuf Adnan YA   Duman Can C  

Biochemia medica 20170828 3


<h4>Introduction</h4>Although serum-providing blood tubes with a barrier are still widely used due to their significant advantages, the use of blood tubes with a barrier to provide plasma is becoming widespread. We compared 22 analytes in a BD Vacutainer® Barricor LH Plasma tube for local clinical validation of this new lithium heparin tube with a barrier.<h4>Materials and methods</h4>Samples from 44 volunteers were collected in different tubes (Becton Dickinson and Company): Z tube without addi  ...[more]

Similar Datasets

| S-EPMC7569594 | biostudies-literature
| S-EPMC6593183 | biostudies-literature
| S-EPMC6938909 | biostudies-literature
| S-EPMC4546686 | biostudies-literature
| S-EPMC7642820 | biostudies-literature
| S-EPMC7570886 | biostudies-literature
| S-EPMC4549105 | biostudies-literature
| S-EPMC6976745 | biostudies-literature
| S-EPMC8566095 | biostudies-literature
| S-EPMC10964274 | biostudies-literature