ABSTRACT: Background
Centrifugation is a consuming time step which participates to increase the turnaround time (TAT) in laboratories, likewise hemolysis sample that needs a re-sampling could delay management of patients. Recently, it has been postulated that BD Barricor™ tube could allow to decrease the centrifugation time and prevent hemolysis, two key feature to ensure high-quality results.Aim of the study was to evaluate the impact of replacing 4?mL BD vacutainer heparin lithium tube by low vacuum 3.5?mL BD vacutainer Barricor™ tube in an emergency department (ED) on hemolysis rate and TAT.Methods
Data of hemolysis index (HI) and TAT were compared between the first period of 15 days using 4?mL BD vacutainer heparin lithium tubes with 15?min?at 2000xg as centrifugation setting and a second period of 15 days using BD vacutainer Barricor™ tube centrifuged 3?min?at 4000xg.Results
A significantly reduced time duration between reception of sample and available results in informatics lab system was observed with the reduction time of centrifugation allowed by use of Barricor™ tube compared to regular heparin lithium tubes (p?ConclusionLow vacuum BarricorTM tubes allowing a higher speed of centrifugation improve lab TAT without impairment of sample quality as a significant reduction of hemolysis was observed, a double advantage which is of particular interest for ED.
SUBMITTER: Badiou S
PROVIDER: S-EPMC6593183 | biostudies-literature | 2019 Aug
REPOSITORIES: biostudies-literature
Badiou S S Vuillot O O Bargnoux A S AS Kuster N N Lefebvre S S Sebbane M M Cristol J P JP Dupuy A M AM
Practical laboratory medicine 20190612
<h4>Background</h4>Centrifugation is a consuming time step which participates to increase the turnaround time (TAT) in laboratories, likewise hemolysis sample that needs a re-sampling could delay management of patients. Recently, it has been postulated that BD Barricor™ tube could allow to decrease the centrifugation time and prevent hemolysis, two key feature to ensure high-quality results.Aim of the study was to evaluate the impact of replacing 4 mL BD vacutainer heparin lithium tube by low va ...[more]