Determination of kQ using MLC-collimated rectangular fields for absolute dosimetry of the CyberKnife.


ABSTRACT: Traditional CyberKnife (CK) calibration uses TG-51, which requires kQ to be defined using the standard reference condition of 100 cm SSD in a 10 cm × 10 cm field. Since the CK is calibrated using a 6 cm fixed-aperture collimating cone at 80 cm SAD, the BJR-25 method is commonly used to relate circular-field PDDs to square-field PDDs for kQ determination. Using the InCise MLC system, the CK is able to deliver rectangular fields, allowing a more direct measurement of %dd(10 cm) using conventional reference conditions. We define the PDD correction factor (CPDD) as the ratio of %dd(10 cm) measured using CK reference conditions to that measured using standard TG-51 reference conditions. Using four ionization chambers (A1SL, CC08, CC13, and A19), %dd(10 cm) is measured using a 6 cm fixed cone at 80 cm SSD and at 100 cm SSD using an effective 10 cm × 10 cm MLC-collimated field. These values are used to calculate CPDD, while the latter is used to directly calculate a kQ value. This direct kQ value is then compared to values determined using the BJR-25 method. Using the MLC system, this study demonstrates conversion between the %dd(10 cm) measured using CyberKnife reference conditions and TG-51 reference conditions. These values provide the means for derivation of a kQ curve as a function of direct measurements of %dd(10 cm) using a 6 cm fixed-aperture collimating cone at 80 cm SSD.

SUBMITTER: Gersh JA 

PROVIDER: S-EPMC5690991 | biostudies-literature | 2015 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications

Determination of kQ using MLC-collimated rectangular fields for absolute dosimetry of the CyberKnife.

Gersh Jacob A JA   Willett Benjamin B  

Journal of applied clinical medical physics 20151108 6


Traditional CyberKnife (CK) calibration uses TG-51, which requires kQ to be defined using the standard reference condition of 100 cm SSD in a 10 cm × 10 cm field. Since the CK is calibrated using a 6 cm fixed-aperture collimating cone at 80 cm SAD, the BJR-25 method is commonly used to relate circular-field PDDs to square-field PDDs for kQ determination. Using the InCise MLC system, the CK is able to deliver rectangular fields, allowing a more direct measurement of %dd(10 cm) using conventional  ...[more]

Similar Datasets

| S-EPMC4926790 | biostudies-literature
| S-EPMC5690184 | biostudies-literature
| S-EPMC3179685 | biostudies-literature
| S-EPMC7241287 | biostudies-literature
| S-EPMC5509254 | biostudies-literature
| S-EPMC5690523 | biostudies-literature
| S-EPMC3702370 | biostudies-literature
| S-EPMC9278669 | biostudies-literature
| S-EPMC6364845 | biostudies-literature
| S-EPMC6025468 | biostudies-literature