Unknown

Dataset Information

Implementation of an improved dose-per-MU model for double-scattered proton beams to address interbeamline modulation width variability.


ABSTRACT: Because treatment planning systems (TPSs) generally do not provide monitor units (MUs) for double-scattered proton plans, models to predict MUs as a function of the range and the nominal modulation width requested of the beam delivery system, such as the one developed by the MGH group, have been proposed. For a given nominal modulation width, however, the measured modulation width depends on the accuracy of the vendor's calibration process and may differ from this nominal value, and also from one beamline to the next. Although such a difference can be replicated in our TPS, the output dependence on range and modulation width for each beam option or suboption has to be modeled separately for each beamline in order to achieve maximal 3% inaccuracy. As a consequence, the MGH output model may

SUBMITTER: Lin L 

PROVIDER: S-EPMC5711055 | biostudies-literature | 2014 May

REPOSITORIES: biostudies-literature