Bead debonding from a modern cementless total hip femoral component with concomitant taper corrosion.


ABSTRACT: We report a case of bead shedding from a cylindrical extensively porous-coated cementless femoral component with concomitant taper corrosion at the modular head-neck junction of a metal-on-polyethylene total hip prosthesis. The patient presented with chronic thigh pain 4 years after primary total hip arthroplasty, and radiographs revealed significant osteolysis and metallic debris around the femoral stem. Intraoperatively, the patient had a grossly loose femoral component with debonding of sintered beads from the femoral stem, as well as evidence of taper corrosion. We identify a failure of a modern beaded femoral component in conjunction with taper corrosion.

SUBMITTER: Edelstein AI 

PROVIDER: S-EPMC5712029 | biostudies-literature | 2017 Dec

REPOSITORIES: biostudies-literature

altmetric image

Publications

Bead debonding from a modern cementless total hip femoral component with concomitant taper corrosion.

Edelstein Adam I AI   Phan Duy L DL   Sporer Scott M SM  

Arthroplasty today 20170506 4


We report a case of bead shedding from a cylindrical extensively porous-coated cementless femoral component with concomitant taper corrosion at the modular head-neck junction of a metal-on-polyethylene total hip prosthesis. The patient presented with chronic thigh pain 4 years after primary total hip arthroplasty, and radiographs revealed significant osteolysis and metallic debris around the femoral stem. Intraoperatively, the patient had a grossly loose femoral component with debonding of sinte  ...[more]

Similar Datasets

| S-EPMC7303489 | biostudies-literature
| S-EPMC7218161 | biostudies-literature
| S-EPMC5264572 | biostudies-literature
| S-EPMC5247516 | biostudies-literature