Comparison of planning techniques when air/fluid is present using the strut-adjusted volume implant (SAVI) for HDR-based accelerated partial breast irradiation.


ABSTRACT: The presence of air/fluid surrounding implantable devices used for partial breast irradiation may significantly impact dose coverage to at-risk tissue. Of the 67 total patients retrospectively evaluated for this study, 32 (48%) had greater than 1 cc volume of air/fluid extending outside of the strut-adjusted volume implant (SAVI) device surface and were selected for comparison of planning approaches. The planning approaches utilized two different definitions of PTV_EVAL. One definition of a PTV_EVAL (PTV_EVALSAVI) was based on expanding 1 cm beyond the SAVI device only while accounting for the air/fluid using the NSABP Protocol B-39/RTOG Protocol 0413. The second PTV_EVAL definition (PTV_EVALCAV) was based on expanding 1 cm beyond the cavity (SAVI device plus air/fluid volume). The results indicate use of the B-39 formalism to account for air/fluid displacing the PTV_EVAL may overestimate the dose coverage to the at-risk tissue, especially for large contiguous volumes of air/fluid. Using the SAVI device to optimize dose covering the PTV_EVALCAV volume surrounding the cavity improves dosimetric coverage to at-risk tissue by 11.3% and 8.7% for V100 and V90, respectively, while the average V150 and V200 indices for PTV_EVALCAV increased by 9.1 cc and 5.0cc, respectively, and the average maximum rib and skin doses increased by 11.1% and 6.1%, respectively. The maximum skin dose, rib dose, V150, and V200 all met the planning objectives despite any increase in these parameters.

SUBMITTER: Harmon JF 

PROVIDER: S-EPMC5714640 | biostudies-literature | 2013 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications

Comparison of planning techniques when air/fluid is present using the strut-adjusted volume implant (SAVI) for HDR-based accelerated partial breast irradiation.

Harmon Joseph F JF   Rice Brandon K BK  

Journal of applied clinical medical physics 20131104 6


The presence of air/fluid surrounding implantable devices used for partial breast irradiation may significantly impact dose coverage to at-risk tissue. Of the 67 total patients retrospectively evaluated for this study, 32 (48%) had greater than 1 cc volume of air/fluid extending outside of the strut-adjusted volume implant (SAVI) device surface and were selected for comparison of planning approaches. The planning approaches utilized two different definitions of PTV_EVAL. One definition of a PTV_  ...[more]

Similar Datasets

| S-EPMC5056328 | biostudies-literature
| S-EPMC7461936 | biostudies-literature
| S-EPMC4608660 | biostudies-literature
| S-EPMC10248091 | biostudies-literature
| S-EPMC9950631 | biostudies-literature
| S-EPMC6136182 | biostudies-literature
| S-EPMC6925849 | biostudies-literature
| S-EPMC6411019 | biostudies-literature
| S-EPMC6380720 | biostudies-literature
| S-EPMC10163674 | biostudies-literature